Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 12 mm 027 REF PED2-300-12
Micro Therapeutics, Inc. began a recall of Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 12 mm 027 REF PED2-300-12 on . FDA classified it as Class II and gives the reason as "Incorrect labeling - embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants contain 20 mm implants". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-2773-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medtronic, Inc.
- Product
- Pipeline Flex Embolization Device with Shield Technology, 3.00 mm x 12 mm 027 REF PED2-300-12
- Product code
- OUT Intracranial Aneurysm Flow Diverter
- Reason for recall
- Incorrect labeling - embolization devices labeled as 3.0 mm x 20 mm implants contain 12 mm implants, and devices labeled as 3.0 mm x 12 mm implants contain 20 mm implants.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- On June 25, 2026, Medtronic (parent company of Micro Therapeutics) issued a Urgent Medical Device Expansion to affected consignees via visit. The revised notification included additional lots of product. The initial notification was issued on June 4, 2026 via visit. Medtronic ask consignees to take the following actions: 1. Do NOT use any impacted Pipeline Shield products. Remove and quarantine all unused impacted product units in your inventory. 2. Return the impacted product(s) to Medtronic. Your Medtronic representative can assist in facilitating the return of product and/or identifying suitable replacement product as necessary. 3. Complete the attached Customer Confirmation Form and email it to [email protected]. 4. Please share this communication within your organization, with other organizations where impacted devices have been transferred, and any other associated organizations that may be impacted by this action. 5. Please maintain a copy of this letter for your records and the records of your patients with Pipeline Shield.
- Quantity in commerce
- 13 units
- Distribution
- US Nationwide distribution in the states of CT, DE, IN, MA, NC, NJ, NM, NY, OH, PA, TX, VA, WA.
- Product codes and lots
- UDI: 00763000448806/ Lot: D062612, D112860
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2773-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2774-2026 | Class II | Ongoing | Voluntary: Firm initiated |