Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC...
AVID Medical, Inc. began a recall of Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC CATH PACK EAMC1000-05 on . FDA classified it as Class I and gives the reason as "Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-2796-2026
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Avid Medical, Inc. a Medical Action Ind. Company 305
- Product
- Halyard convenience kits containing recalled Medline Namic Angiographic Star Off Handle Manifolds. CARDIAC CATH PACK EAMC1000-05
- Product code
- OES Cardiac Catheterization Kit
- Reason for recall
- Affected kits contain Medline Namic Angiographic Star Off Handle Manifolds, which were recalled by the manufacturer due to presence of particulates within fluid path of the Manifolds. Particulate within the fluid path has the potential to be introduced into blood circulation and become lodged within blood vessels.
- Root cause tag
- manufacturing process
- FDA root cause
- Under Investigation by firm
- Action
- On April 30, 2026, firm notified distributors via email. Distributors were tasked with notifying their end-users. Customers were instructed to identify, segregate, and quarantine all impacted products. They were provided with a labeling template to print warning labels which should be affixed on all affected kits and cases in inventory. The firm is also able to provide pre-printed labels if needed. Warning labels inform users to remove and discard the manifold prior to use.
- Quantity in commerce
- 5168 kits
- Distribution
- US Nationwide distribution in the states of NC and TX.
- Product codes and lots
- UDI-DI 10809160377314 Lots 1573593 1583827 1609541 1612134 1612516 1619132 1623384 1627013 1635459 1672306 1675159
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2796-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2797-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2798-2026 | Class I | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY