Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-2867-2026Class II

Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary Bone...

Zimmer GmbH / initiated 2026-07-01 / Open, Classified / root cause: packaging

Biomet, Inc. began a recall of Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary Bone Saw Blade Assembly 12mm on . FDA classified it as Class II and gives the reason as "Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Saw Blade Assembly due to potential sterile barrier packaging damage caused by an exposed saw blade". As of , the recall status on record is Open, Classified, and MDM tags the root cause as packaging.

Recall record

Recall number
Z-2867-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Zimmer GmbH
Product
Intramedullary Bone Saw Blade Assembly Model/Catalog Number: 475620 Product Description: Intramedullary Bone Saw Blade Assembly 12mm
Product code
LXH Orthopedic Manual Surgical Instrument
Reason for recall
Zimmer Biomet is conducting a medical device recall removal for various sizes of the Intramedullary Bone Saw Blade Assembly due to potential sterile barrier packaging damage caused by an exposed saw blade.
Root cause tag
packaging
FDA root cause
Under Investigation by firm
Action
An URGENT MEDICAL DEVICE RECALL notification letter dated 6/22/26 was sent to customers. Risk Manager Responsibilities: 1. Review this notification and ensure that affected personnel are aware of the contents. 2. If you have affected product at your facility, assist your Zimmer Biomet sales representative and quarantine all affected product. Your Zimmer Biomet sales representative will remove the affected product from your facility. 3. If the product has been further distributed, provide your customers with this recall notice and ensure documentation. 4. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. This form shall be returned even if you do not have affected products at your facility. 5. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 6. If you have further questions or concerns after reviewing this notice, please call customer service at 1-800-348-2759 between 8:00 a.m. and 5:00 p.m. Eastern, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail. Alternatively, your questions may be emailed to [email protected]. Surgeon Responsibilities: 1. Review this notification for awareness of the contents. 2. There are no specific patient monitoring instructions related to this recall that are recommended beyond your existing follow-up schedule. 3. Complete Attachment 1 Certificate of Acknowledgement and send to [email protected]. 4. Retain a copy of the acknowledgement form with your recall records in the event of a compliance audit of your facility s documentation. 5. If you have further questions or concerns after reviewing this notice, please call customer service at 1-800-348-2759 between 8:00 a.m. and 5:00 p.m. Eastern, Monday through Friday. Calls received outside of call center operating hours will receive a vo
Quantity in commerce
10
Distribution
Refer to the attached ZFA 2026-00096_Initial Consignee List and ZFA 2026-00096_Initial Distribution History containing the distribution and the identification of the affected consignees.
Product codes and lots
UDI-DI (01)00887868507793(17)360407(10)67818791; Lot 67818791

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-2867-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-2867-2026Class IIOngoingVoluntary: Firm initiated
Z-2868-2026Class IIOngoingVoluntary: Firm initiated
Z-2869-2026Class IIOngoingVoluntary: Firm initiated
Z-2870-2026Class IIOngoingVoluntary: Firm initiated
Z-2871-2026Class IIOngoingVoluntary: Firm initiated