Brilliance iCT Brilliance iCT Upgrades Brilliance iCT SP Brilliance CT 64 Channel CT6000 Brilliance 64...
PHILIPS MEDICAL SYSTEMS began a recall of Brilliance iCT Brilliance iCT Upgrades Brilliance iCT SP Brilliance CT 64 Channel CT6000 Brilliance 64 Upgrades on . FDA classified it as Class II and gives the reason as "The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being...". As of , the recall status on record is Open, Classified.
Recall record
- Recall number
- Z-2650-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- Brilliance iCT Brilliance iCT Upgrades Brilliance iCT SP Brilliance CT 64 Channel CT6000 Brilliance 64 Upgrades
- Product code
- JAK System, X-Ray, Tomography, Computed
- Reason for recall
- The clinical CT scan may not automatically trigger after the Bolus Tracking scan and may need to be triggered manually. Additional Bolus Tracking shots may occur. This may result in the clinical acquisition being inadequate for diagnosis, causing a potential for patient rescan and receiving additional radiation.
- FDA root cause
- Radiation Control for Health and Safety Act
- Action
- An Electronic Product Radiation Correction letter was issued to its customers describing the product, problem and actions to be taken by its customers. To identify the software version of your product: 1. Click the Help button. 2. Select About and the software version is then displayed. The software version begins with v. 3.You may continue to use your system(s) in accordance with the intended use. 4. Please complete and return the attached response form to Philips promptly and no later than 30 days from receipt. Completing this form confirms receipt of the Electronic Product Radiation Correction Letter, understanding of the issues, and required actions to be taken. 5. Circulate this Electronic Product Radiation Correction Letter to all users of this device so that they are aware of the issues. Please retain this letter with your system(s) until a solution is installed on your system; ensure the letter is in a place likely to be seen/viewed. Actions planned by Philips to correct the problem Philips will contact you to schedule a time for a Philips Field Service Engineer (FSE) to visit your site and install the software update to resolve the issues. If you need any further information or support concerning this issue, please contact your local Philips representative: For North America, contact the Customer Care Solutions Center (1-800-722-9377).
- Quantity in commerce
- 128
- Distribution
- US Nationwide Distribution including Puerto Rico
- Product codes and lots
- FCO72800812; FCO72800814 728305 728306 DI (01)00884838059474 728307 DI (01)00884838104600 728311 728231 (01)00884838083325 728232
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2650-2026 | Class II | Ongoing | FDA Mandated | ||
| Z-2651-2026 | Class II | Ongoing | FDA Mandated | ||
| Z-2652-2026 | Class II | Ongoing | FDA Mandated |