Z-2955-2026Class II
Cardiac Workstation 5000. Product Number: 860439
Philips North America LLC / initiated 2026-07-01 / Open, Classified / root cause: software
Philips North America Llc began a recall of Cardiac Workstation 5000. Product Number: 860439 on . FDA classified it as Class II and gives the reason as "Device will intermittently provide inaccurate data related to native heart rate". As of , the recall status on record is Open, Classified, and MDM tags the root cause as software.
Recall record
- Recall number
- Z-2955-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America LLC
- Product
- Cardiac Workstation 5000. Product Number: 860439.
- Product code
- DPS Electrocardiograph
- Reason for recall
- Device will intermittently provide inaccurate data related to native heart rate.
- Root cause tag
- software
- FDA root cause
- Software design
- Action
- Philips notified consignees on about 07/01/2026 via URGENT Medical Device Correction letter. Consignees were informed that they could continue the use of the device but to refer to the information within the notification letter, disable the automatic Standby feature on the Cardiac Workstation and avoid manually putting the Cardiac Workstation in Standby mode to minimize the inaccurate heart rate value from occurring, and to pass the notice to all those who need to be aware within each organization or to any organization where affected product(s) have been potentially transferred. Additionally, consignees were requested to place the Urgent Medical Device Correction letter with the documentation of the Cardiac Workstation and associated devices in a place where it is most likely to be seen and viewed. Philips representatives will contact consignees to arrange an upgrade to the Cardiac Workstation Release 1.2.0.7.
- Quantity in commerce
- 796 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia, Austria, Bahrain, Belgium, Brazil, Canada, Croatia, Czech Republic, Dominican Republic, France, Germany, Gibraltar, Hong Kong, Iceland, Ireland, Iraq, Italy, Luxembourg, Malta, Netherlands, New Caledonia, New Zealand, Norway, Poland, Portugal, Qatar, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Korea, Spain, Sweden, Switzerland, Taiwan, Thailand, United Arab Emirates, United Kingdom.
- Product codes and lots
- Product Number: 860439. UDI: (01)00884838094826. Serial Numbers: US42510185, US42510184, US52510209, US52510200, US52510226, US52510228, US62510247, US62510250, US62510256, US72510286, US82510299, US82510307, US82510300, US82510313, US82510318, US82510314, US82510335, US82510336, US82510341, US82510325, US92510351, US92510353, US92510354, US92510361, US92510362, US92510370, USN2510413, USN2510414, USN2510434, USN2510435, USN2510437, USD2510506, USD2510519, US12610582, USD2510513, US12610598, US22610609, US22610683, US22610691, US22610689, US32610737, US32610744, US32610759, US32610766, US32610749, US32610739, US32610751, USD2510516, USD2510510, USD2510517, USD2510529, USD2510509, US32610752, US32610753, US32610879, US32610880, US32610901, US32610914, US42610946, US32610915, US32610941, US32610932, US42610943, US42610942, US42610959, US42610958, US42610955, US42610974, US42610967, US42610956, US42610993, US42610960, CN72410032, CN92410061, CNO2410064, CNO2410068, CNN2410075, CND2410080, CND2410101, CND2410103, CN12510116, CN12510136, CN22510140, CN22510142, CN22510146, CN32510148, CN32510161, CN42510168, CN42510181, CN42510183, CN52510204, CN52510201, CN52510217, CN62510239, CN62510241, CN62510242, CN62510253, CN62510254, CN62510251, CN72510275, CN72510282, CN72510281, CN72510285, CN72510289, CN72510288, CN72510293, CN72510297, CN82510305, CN82510310, CN82510323, CN82510324, CN82510321, CN92510343, CN92510364, CN92510365, CN92510367, CN92510374, CNO2510380, CNO2510392, CNO2510383, CNO2510386, CNO2510396, CNN2510412, CNN2510410, CNN2510427, CNN2510443, CNN2510444, CNN2510454, CNN2510485, CNN2510486, CND2510503, CND2510502, CND2510508, CND2510528, CND2510530, CND2510566, CND2510575, CND2510574, CN12610585, CN12610590, CN12610595, CN12610593, CN22610612, CN22610681, CN22610707, CN22610735, CN32610755, CN32610760, CN32610761, CN32610762, CN32610763, CN32610765, CN32610818, CN32610826, CN32610910, CN32610913, CN32610912, CN32610924, CN32610923, CN32610929, CN32610930, CN32610933, CN42610944, CN42610951, CN42610970, US42610994
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2955-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2956-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.