Z-0076-2027
Brand Name: ADAPTABLE Beach Chair Product Name: Beach Chair surgical patient positioner Model/Catalog Number...
QUANTUM OPS Inc / initiated 2026-07-24 / Open, Classified
Recall record
- Recall number
- Z-0076-2027
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- QUANTUM OPS Inc
- Product
- Brand Name: ADAPTABLE Beach Chair Product Name: Beach Chair surgical patient positioner Model/Catalog Number: 804-33-300 (Enovis); 3105-400-030 (Stryker); internal BC-0001 / BC-0220 Software Version: N/A Product Description: Surgical beach-chair positioner that mounts to the accessory rails of the OR table leg section to elevate and support the patient's upper body in the upright (beach chair) position during surgical procedures. Component: No, finished device. The failure is in the BC-0105, subcomponent.
- Product code
- BWN Table And Attachments, Operating-Room
- Reason for recall
- Due to a manufacturing issue (inadequate weld) for patient head positioning system.
- FDA root cause
- Device Design
- Action
- On 07/24 and 07/27/2026, an "URGENT: Medical Device Recall" letter was emailed to customers informing them that, the firm has received complaints describing the fracture of a welded joint below the head rest which provides support to patient s head. Customers are instructed to: 1. Inform individuals within organization who need to be aware, and pass this notice on to any organization where the product has been transferred. 2. Check internal inventory to locate the affected product. 3. Perform the device safety check as detailed in Attachment B and complete the Acknowledgement and Receipt Form (Attachment A) and return to [email protected] or [email protected] 4. If the device passes, it is suitable to continue clinical use. It will be scheduled for exchange with an upgraded unit. a. Upon completing the Acknowledgement and Receipt Form (Attachment A), contact Enovis Customer Service at 800-520-8976 or email [email protected] to schedule the exchange, or b. Upon completing the Business Reply Form (Attachment A), Stryker will contact the account to schedule the exchange. 5. If the device fails, segregate the device and do not continue use. a. Upon completing the Acknowledgement and Receipt Form (Attachment A), please contact Enovis Customer Service at 800-520-8976 or email [email protected] to arrange for return and issuance of replacement (upon availability), or b. Upon completing the Business Reply Form (Attachment A), Stryker will contact your account to arrange for return and issuance of replacement (upon availability). -The returned device will be replaced with an upgraded unit. 6. If affected product is NOT found: a. Please complete Acknowledgement and Receipt Form (Attachment A). 8. Maintain awareness of this communication internally until all required actions have been completed within their facility. Response is required.
- Quantity in commerce
- 266
- Distribution
- U.S. Nationwide distribution in the states of CA and TX.
- Product codes and lots
- Lot Code: GTIN / UDI-DI: 00190446667278 Affected lots / serials: Product Number: 3105-400-030 Lots: 2021110101 Serial Numbers: A-026 A-027 A-028 A-029 A-030 A-031 A-032 A-033 A-034 A-035 A-036 A-037 A-038 A-039 A-040 Lot Number: 2021120201 Serial Numbers: A-041 A-042 A-043 A-044 A-045 A-046 A-047 A-048 A-049 A-050 A-051 A-052 A-053 A-054 A-056 A-057 A-058 A-059 A-060 A-061 A-062 A-063 A-064 A-065 A-066 A-067 A-068 A-069 A-070 Lot Number: 2022010601 Serial Numbers: A-071 A-072 A-073 A-074 A-075 A-076 A-077 A-078 A-079 A-082 A-083 A-084 A-087 A-090 A-091 A-092 A-093 A-094 A-095 A-099 Lot Number: 2023012501 Serial Numbers: C-051 C-052 C-053 C-054 C-055 C-056 C-057 C-058 C-059 C-060 C-061 C-062 C-063 C-064 C-065 C-066 C-067 Lot Number: 2023020201 Serial Numbers: C-001 C-002 C-003 C-004 C-005 C-006 C-007 C-008 C-009 C-010 C-011 C-012 C-013 C-014 C-015 C-016 C-017 C-018 C-019 C-020 C-021 C-022 C-023 C-024 C-025 C-026 C-027 C-028 C-029 C-030 C-031 Lot Number: 2023060101 Serial Numbers: C-081 C-082 C-083 C-084 C-085 C-086 C-087 C-088 C-089 C-090 C-091 C-092 C-093 C-094 C-095 Lot Number: 2023071801 Serial Numbers: C-096 C-097 C-098 C-099 C-100 C-101 C-102 C-103 C-104 C-105 C-106 C-107 C-108 C-109 C-110 C-111 C-112 C-113 C-114 C-115 C-116 C-117 C-118 C-119 C-120 C-121 C-122 Lot Number: 2024081401 Serial Numbers: C-123 C-124 C-125 C-126 C-127 C-128 C-129 C-130 C-131 C-132 Lot Number: 2024091001 Serial Numbers: C-133 C-134 C-135 C-136 C-137 C-138 C-139 C-140 C-141 C-142 C-143 C-144 C-145 C-146 C-147 Lot Number: 2024100301 Serial Numbers: C-148 C-149 C-150 C-151 C-152 C-153 C-154 C-155 C-156 C-157 Lot Number: 2024110401 Serial Numbers: C-158 C-159 C-160 C-161 C-162 C-163 C-164 C-165 C-166 C-167 C-168 C-169 C-170 C-171 C-172 Product Number: 804-33-300 Lot Number: 2020102701 Serial Numbers: A-001 A-002 A-003 A-004 A-005 A-006 A-007 A-010 A-011 A-012 Lot Number: 2021050601 Serial Numbers: A-014 A-015 A-016 A-017 A-018 A-019 A-020 A-021 A-023 A-024 Lot Number: 2022031701 Serial Numbers: A-0131 A-0132 A-0133 A-0134 A-0135 A-0136 A-0138 A-0139 A-0140 Lot Number: 2023020101 Serial Numbers: C-068 C-069 C-070 C-073 C-071 C-072 C-074 C-075 C-076 C-077 C-078 C-079 C-080 Lot Number: 2023032001 Serial Numbers: C-032 C-033 C-034 C-035 C-036 C-037 C-038 C-039 C-040 C-041 C-042 C-043 C-044 C-045 C-046 C-047 C-048 C-049 C-050 2020101201. Excluded (demo/undelivered): A-008, A-009, A-013, A-022, A-025, A-0137.