TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798
Tobii Dynavox Llc began a recall of TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798 on . FDA classified it as Class II and gives the reason as "Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2959-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Tobii Dynavox LLC
- Product
- TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798
- Product code
- ILQ System, Communication, Powered
- Reason for recall
- Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- On July 8, 2026, the firm notified customers by emailing letters titled "URGENT: MEDICAL DEVICE RECALL NOTICE." Customers were informed that Tobii Dynavox would provide a repair authorization number (RA#), prepaid shipping label, inspection, repair, and return shipping for the affected device. The repair will include modifications intended to improve battery stability within the device. Customers were provided with ways to contact the company to obtain a repair authorization number. Customers were informed that it is safe to continue using the device as intended while waiting for the repair, as long as there are no signs of overheating, swelling, unusual odor, smoke, deformation, or battery abnormality. If any of these conditions are observed, discontinue use immediately and contact Tobbi Dynavox.
- Distribution
- US Nationwide distribution.
- Product codes and lots
- UDI-DI 073400747602056 Serial Numbers TDNM1-240628000004 through TDNM1-260617070001. Corrected/future production is differentiated by new serialization from TDNM1-260617070001 onward (production batch 20260615, improved design).
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2959-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2960-2026 | Class II | Ongoing | Voluntary: Firm initiated |