Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3046-2026Class II

Halyard medical convenience kits labeled as: 1) A-P PACK - 205936, Model/Item Number (kit) ANAP22-01; 2)...

Avid Medical, Inc. a Medical Action Ind. Company 305 / initiated 2026-05-19 / Open, Classified / root cause: manufacturing process

AVID Medical, Inc. began a recall of Halyard medical convenience kits labeled as: 1) A-P PACK - 205936, Model/Item Number (kit) ANAP22-01; 2) POSTERIOR PACK - 205960, Model/Item Number (kit)... on . The reason FDA records is "The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3046-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Avid Medical, Inc. a Medical Action Ind. Company 305
Product
Halyard medical convenience kits labeled as: 1) A-P PACK - 205936, Model/Item Number (kit) ANAP22-01; 2) POSTERIOR PACK - 205960, Model/Item Number (kit) ANPP15-01; 3) ANEURYSM PACK, Model/Item Number (kit) BPAY14-01; 4) SPINE PACK, Model/Item Number (kit) HSLM53-01; 5) ANTERIOR CERVICAL PACK - 205988, Model/Item Number (kit) MHAC65-01; 6) SPINE PACK - 206013, Model/Item Number (kit) MHSP99-01; 7) CRANI PACK, Model/Item Number (kit) WECR41-01.
Product code
OJH Orthopedic Tray
Reason for recall
The recall notice was initiated due to a Medicom Urgent Medical Device Recall. Medicom is recalling specific lots of Ritmed Neurological Sponges Non-Sterile Models 55992C to 55998C due to potential for unexpected variability of endotoxin levels on the products.
Root cause tag
manufacturing process
FDA root cause
Material/Component Contamination
Action
Owens & Minor issued an URGENT: MEDICAL DEVICE RECALL NOTIFICATION to its consignees on 5/19/2026 via email. The notice explained the issue, potential risk to health and requested the following: "ACTIONS TO TAKE: " Identify, segregate, and quarantine all impacted product. " Add recall labels to affected Avid kits to ensure all users are aware of the actions to discard the impacted component. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. Please contact your Sales Representative if pre-printed labels are needed. This label should be applied in a prominently visible location to end-users. The location chosen should not cover any other critical product information found on existing product labeling. " Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. " Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness."
Quantity in commerce
816 kits
Distribution
US Nationwide distribution in the states of AZ, CA, CO, WA, WY, TX, MI, FL, MD, UT, NC, MS, PA, MO, IA, NE, OH, IL, SD, MN, GA.
Product codes and lots
Model/Item Number (kit) ANAP22-01, UDI/DI 10809160473146, Kit Lot Number 346967; Model/Item Number (kit) ANAP22-01, UDI/DI 10809160473146, Kit Lot Number 346982; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343413; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343414; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343713; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343715; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343716; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343412; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343416; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343714; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343717; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343718; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343719; Model/Item Number (kit) ANPP15-01, UDI/DI 10809160473573, Kit Lot Number 343723; Model/Item Number (kit) BPAY14-01, UDI/DI 10809160473979, Kit Lot Number 343746; Model/Item Number (kit) BPAY14-01, UDI/DI 10809160473979, Kit Lot Number 343747; Model/Item Number (kit) BPAY14-01, UDI/DI 10809160473979, Kit Lot Number 343771; Model/Item Number (kit) BPAY14-01, UDI/DI 10809160473979, Kit Lot Number 374128; Model/Item Number (kit) BPAY14-01, UDI/DI 10809160473979, Kit Lot Number 374129; Model/Item Number (kit) BPAY14-01, UDI/DI 10809160473979, Kit Lot Number 374296; Model/Item Number (kit) HSLM53-01, UDI/DI 10809160476888, Kit Lot Number 342005; Model/Item Number (kit) HSLM53-01, UDI/DI 10809160476888, Kit Lot Number 344071; Model/Item Number (kit) HSLM53-01, UDI/DI 10809160476888, Kit Lot Number 347906; Model/Item Number (kit) HSLM53-01, UDI/DI 10809160476888, Kit Lot Number 341835; Model/Item Number (kit) HSLM53-01, UDI/DI 10809160476888, Kit Lot Number 343206; Model/Item Number (kit) HSLM53-01, UDI/DI 10809160476888, Kit Lot Number 343208; Model/Item Number (kit) MHAC65-01, UDI/DI 10809160479537, Kit Lot Number 345289; Model/Item Number (kit) MHSP99-01, UDI/DI 10809160479797, Kit Lot Number 345694; Model/Item Number (kit) MHSP99-01, UDI/DI 10809160479797, Kit Lot Number 341654; Model/Item Number (kit) WECR41-01, UDI/DI 10809160485392, Kit Lot Number 344464.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3046-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3044-2026Class IIOngoingVoluntary: Firm initiated
Z-3045-2026Class IIOngoingVoluntary: Firm initiated
Z-3046-2026Class IIOngoingVoluntary: Firm initiated