Z-2968-2026Class II
Blanketrol II Model 222S; Thermal regulating system
Gentherm Medical, LLC / initiated 2026-07-13 / Open, Classified / root cause: design
Gentherm Medical, LLC began a recall of Blanketrol II Model 222S; Thermal regulating system on . FDA classified it as Class II and gives the reason as "Overheating occurring on 115V CPU boards for the Blanketrol III Model 233, Innercool CoolBlue Model STx, and Blanketrol II Model 222S products may cause the device to shut down". As of , the recall status on record is Open, Classified, and MDM tags the root cause as design.
Recall record
- Recall number
- Z-2968-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Gentherm Medical, LLC
- Product
- Blanketrol II Model 222S; Thermal regulating system.
- Product code
- DWJ System, Thermal Regulating
- Reason for recall
- Overheating occurring on 115V CPU boards for the Blanketrol III Model 233, Innercool CoolBlue Model STx, and Blanketrol II Model 222S products may cause the device to shut down.
- Root cause tag
- design
- FDA root cause
- Device Design
- Action
- Genthem issued an URGENT Medical Device Notice to its consignees on 7/17/2026 via email. The notice explained the issue, potential risk, and request the following: "Interim Recommendations: 1. Based on the low probability of harm, and absence of reported harm, associated with this issue, users may continue to use the device. In an abundance of caution, it is recommended to have an alternative means of providing therapy available. 2. Users should review and follow the device Instructions for Use; users should routinely verify the device s function and patient s condition during use. 3. Ensure that all users are informed of the contents of this letter. If you have further distributed this product, please provide those accounts with a copy of this notice." For questions or if you have experienced the described issue or a related adverse event, call Gentherm Medical Technical Support at 1-888-437-5608.
- Quantity in commerce
- 182 units
- Distribution
- Worldwide - US Nationwide distribution including in the territories of Puerto Rico, and the countries of Brazil, Canada, Chile, Colombia, Costa Rica, El Salvador, Guatemala, Mexico, Republic of Panama, Saudi Arabia, Taiwan, Thailand, Bahamas, and Trinidad and Tobago.
- Product codes and lots
- UDI: 1061303186165 Serial Numbers:131-2-00580 through 262-3-20277
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2968-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2969-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-2970-2026 | Class II | Ongoing | Voluntary: Firm initiated |