Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product...

Greiner Bio-One North America, Inc. / initiated 2026-07-29 / Open, Classified

Recall record

Recall number
Z-0200-2027
Status
Open, Classified
Initiated
Posted
Recalling firm
Greiner Bio-One North America, Inc.
Product
Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml)
Product code
JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
Reason for recall
Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
FDA root cause
Nonconforming Material/Component
Action
On 07/29/2026, an "URGENT MEDICAL DEVICE RECALL NOTICE" was emailed to customers informing them that, Greiner Bio-One North America identified a potential ethylenediaminetetraacetic acid (EDTA) contamination issue in a specific lot of VACUETTE Lithium Heparin serum separator tubes. Use of affected tubes may lead to: " Falsely elevated potassium (K+) results " Falsely decreased calcium (Ca++) results Customers are instructed to: " Immediately identify and quarantine all remaining units of item 454008P, lot B26043FM. " Stop using the affected lot for blood collection. " Notify relevant staff (phlebotomy, lab, clinical teams) of this advisory. " Review any recent patient results obtained using this lot. Consider repeat testing with a different lot if results appear inconsistent with clinical presentation. " Follow proper order of draw, and recollection procedures when re-testing. For unused product disposal (if destroyed by customer/distributor) " The unused blood collection tubes contain chemical additives (LH and serum separator). Recommend following OSHA process, and/or your facility's environmental health and safety (EHS) policies and procedures for disposal. For questions or documentation (Acknowledgement and Receipt Form) to [email protected] or please call 800-515-8112.
Quantity in commerce
860450
Distribution
U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.
Product codes and lots
Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12