Z-0200-2027
Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product...
Greiner Bio-One North America, Inc. / initiated 2026-07-29 / Open, Classified
Recall record
- Recall number
- Z-0200-2027
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Greiner Bio-One North America, Inc.
- Product
- Brand Name: VACUETTE Product Name: LH Lithium Heparin Separator Model/Catalog Number: 454008P Product Description: Greiner VACUETTE Blood Collection Tube with Lithium Heparin and Gel Separator (13 x 75 mm, 3.5ml)
- Product code
- JKA Tubes, Vials, Systems, Serum Separators, Blood Collection
- Reason for recall
- Lot B26043FM may include EDTA contamination which introduces potassium while chelating calcium in patient samples, leading to falsely elevated potassium (K+) and falsely decreased calcium (Ca++) results which have the potential to cause hyperkalemia.
- FDA root cause
- Nonconforming Material/Component
- Action
- On 07/29/2026, an "URGENT MEDICAL DEVICE RECALL NOTICE" was emailed to customers informing them that, Greiner Bio-One North America identified a potential ethylenediaminetetraacetic acid (EDTA) contamination issue in a specific lot of VACUETTE Lithium Heparin serum separator tubes. Use of affected tubes may lead to: " Falsely elevated potassium (K+) results " Falsely decreased calcium (Ca++) results Customers are instructed to: " Immediately identify and quarantine all remaining units of item 454008P, lot B26043FM. " Stop using the affected lot for blood collection. " Notify relevant staff (phlebotomy, lab, clinical teams) of this advisory. " Review any recent patient results obtained using this lot. Consider repeat testing with a different lot if results appear inconsistent with clinical presentation. " Follow proper order of draw, and recollection procedures when re-testing. For unused product disposal (if destroyed by customer/distributor) " The unused blood collection tubes contain chemical additives (LH and serum separator). Recommend following OSHA process, and/or your facility's environmental health and safety (EHS) policies and procedures for disposal. For questions or documentation (Acknowledgement and Receipt Form) to [email protected] or please call 800-515-8112.
- Quantity in commerce
- 860450
- Distribution
- U.S. Nationwide distribution in the states of CO, FL, IL, MO, NE, NY, OH, PA, and VA.
- Product codes and lots
- Lot Code: Item Number: 454008P Lot Number: B26043FM UDI: Rack: 29120017579032, Case: 39120017579039 Expiration Date: 2027-07-12