Z-3009-2026Class II
Halyard TOTAL JOINT Model/Catalog Number: MSBM036-03
Avid Medical, Inc. a Medical Action Ind. Company 305 / initiated 2026-07-09 / Open, Classified / root cause: labeling and UDI
AVID Medical, Inc. began a recall of Halyard TOTAL JOINT Model/Catalog Number: MSBM036-03 on . FDA classified it as Class II and gives the reason as "Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural rubber latex". As of , the recall status on record is Open, Classified, and MDM tags the root cause as labeling and UDI.
Recall record
- Recall number
- Z-3009-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Avid Medical, Inc. a Medical Action Ind. Company 305
- Product
- Halyard TOTAL JOINT Model/Catalog Number: MSBM036-03
- Product code
- LRO General Surgery Tray
- Reason for recall
- Labeling Deficiency: The medical kit labeling fails to identify that the statlock component contains natural rubber latex.
- Root cause tag
- labeling and UDI
- FDA root cause
- Labeling Change Control
- Action
- On July 9, 2026 Avid Medical ( subsidiary of Owen & Minor) issued a Urgent: Medical Device correction notification to affected consignees. Avid Medical ask consignees to take the following actions: 1. Identify, segregate, and quarantine all impacted product identified. 2. Add warning labels to affected AVID kits to ensure all users are aware of the actions to discard the impacted component prior to use. A labeling template is enclosed for printing labels which can be affixed on all affected kits and cases in inventory. 3. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 4. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified and maintain records of effectiveness. 5. Complete the attached Recall Response Form and return by July 24, 2026.
- Quantity in commerce
- 120 Kits
- Distribution
- US Nationwide distribution in the states of MS, TX.
- Product codes and lots
- Lot Code: Lot Number: 1662335 UDI: 10809160457887
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3005-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3006-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3007-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3008-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3009-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3010-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3011-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.