Z-3329-2026
Olympus Tripp-Lite P580-025 25ft DisplayPort Cable, Model Number P580-025, an accessory used in conjunction...
Olympus Corporation of the Americas / initiated 2026-08-27 / Open, Classified
Recall record
- Recall number
- Z-3329-2026
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Olympus Corporation of the Americas
- Product
- Olympus Tripp-Lite P580-025 25ft DisplayPort Cable, Model Number P580-025, an accessory used in conjunction wtih OLYSENSE Hub System
- Product code
- N/A
- Reason for recall
- While in the high-power operating mode, a video dropout/blackout may occur when using a particular accessory cable.
- FDA root cause
- Under Investigation by firm
- Action
- Consignees were notified on about 08/27/2026 via letter. Consignees were instructed to discontinue using the affected cable, arrange for the replacement of the affected cable, acknowledge receipt of the recall notification through their recall web portal, and forward the notification to other users who may have the affected products, if further transferred.
- Quantity in commerce
- 543 cables
- Distribution
- Worldwide distribution - US Nationwide and the country of Poland.
- Product codes and lots
- Model Number P580-025, Material Number Material: OS-DPDPC25-G