Z-2984-2026Class I
Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605150; Power-Trialysis Short-Term Straight...
Bard Access Systems, Inc. / initiated 2026-08-10 / Open, Classified
Recall record
- Recall number
- Z-2984-2026
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Bard Access Systems, Inc.
- Product
- Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605150; Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605200; Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605240; Power-Trialysis Short-Term Straight Dialysis Catheter, REF: 5605300; Power-Trialysis Short-Term Curved Extension Dialysis Catheter, REF: 5615120; Power-Trialysis Short-Term Curved Extension Dialysis Catheter, REF: 5615150; Power-Trialysis Short-Term Curved Extension Dialysis Catheter, REF: 5615200; Power-Trialysis Short-Term Curved Extension Dialysis Catheter, REF: 5615240; Power-Trialysis Short-Term Alphacurve Dialysis Catheter, REF: 5655120; Power-Trialysis Short-Term Alphacurve Dialysis Catheter, REF: 5655150; Power-Trialysis Short-Term Alphacurve Dialysis Catheter, REF: 5655200; Power-Trialysis Short-Term Alphacurve Dialysis Catheter, REF: 5655240; Power-Trialysis Short-Term Straight Dialysis Catheter Full Procedure Tray, REF: 5605200G; Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter, REF: 5855200G; Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter, REF: 5865150G; Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter, REF: 5855150; Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter, REF: 5855200; Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter, REF: 5855240; Power-Trialysis Slim-Cath Straight Short-Term Dialysis Catheter, REF: 5855300; Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter, REF: 5865120; Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter, REF: 5865150; Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter, REF: 5865200; Power-Trialysis Slim-Cath Precurved Extension Short-Term Dialysis Catheter, REF: 5865240
- Product code
- MPB Catheter, Hemodialysis, Non-Implanted
- Reason for recall
- this is a downstream recall of class I recall RES 98777
- FDA root cause
- Material/Component Contamination
- Action
- On 8/10/2026, recall notices were mailed or emailed to customers and distributors who were asked to do the following: 1) Do not use the Sodium chloride ampules contained within the affected products. The sodium chloride ampule is the only impacted component within the tray and all remaining components can continue to be used. 2) Secure an alternative local Sodium chloride prior to entering the procedural environment. 3) Upon opening the tray, remove and destroy the ampules within a controlled and sterile environment to maintain the sterility of the remaining components within the tray. 4) If affected Sodium chloride ampule was previously used without issue, no additional treatment or clinical follow up is required. 5) If affected Sodium chloride ampule was previously used with issue, report the incident per institutional guidelines and take any necessary actions. Take the Following Actions: a) Circulate this notice to all those who need to be aware within your organization or to any organization where the potentially affected products have been transferred. b) Post this notice on all inventory storage locations and any location where the product may be used. c) Label all affected trays using the template provided to facilitate easy identification of products containing the affected ampules. The label can be printed using Avery template 5160. d) Apply the labels to the product without covering or obscuring any regulatory or traceability information. e) Apply to the top web of the tray by pressing firmly to ensure adhesion, but don't damage or compromise the sterile barrier. Distributors are asked to identify all customers within their distribution network that purchased any affected product and to provide a copy of the customer letter to all customers. Complete and return response form via email to [email protected] Report complaints to firm via: Phone: 1-844-823-5433 or Email: [email protected]
- Quantity in commerce
- 149521
- Distribution
- US: FL, IL, OH, VA, MD, TX, WI, GA, TN, IN, SC, NC, MI, PA, OR, OK, AZ, MS, NJ, AL, CA, CO, MA, NV, ID, MO, LA, WA, KY, IA, NH, NY, KS, HI, NE, MN, MT, WY, AR, RI, WV, UT, PR, DE, CT, ME, DC, ND, NM, SD
- Product codes and lots
- REF/UDI-DI/Lot(Expiration): 5605150/801741086953/ REKQ2236(08/31/2026), REKR0948(08/31/2026), REKR3422(09/30/2026), REKR3433(09/30/2026), REKS1563(10/31/2026), REKS3564(09/30/2026), REKS3559(10/31/2026), REKT0573(11/30/2026), REKT1722(09/30/2026), REKT3057(10/31/2026), REKU1678(12/31/2026), REKV0926(01/31/2027), REKW0664(02/28/2027), REKX0410(03/31/2027), REKX1755(03/31/2027), REKY2464(03/31/2027), REKZ2323(05/31/2027), RELN2937(06/30/2027); 5605200/801741086960/ REKQ3355(8/31/2026, REKQ3362(8/31/2026), REKR0756(9/30/2026), REKR2159(9/30/2026), REKR2214(9/30/2026), REKS1550(10/31/2026), REKS1559(10/31/2026), REKS2495(10/31/2026), REKT0576(10/31/2026), REKT0579(11/30/2026), REKT2201(11/30/2026), REKT3018(11/30/2026), REKU2636(12/31/2026), REKU3679(12/31/2026), REKV2415(12/31/2026), REKW2742(2/28/2027), REKX0576(3/31/2027), REKX2423(3/31/2027), REKZ0519(5/31/2027), RELN1850(6/30/2027); 5605240/801741086977/ REKQ3367(8/31/2026), REKR0762(8/31/2026), REKS3554(8/31/2026), REKT2178(11/30/2026), REKU3687(12/31/2026), REKW0663(2/28/2027), REKW2120(2/28/2027), REKX0561(12/31/2026), REKY1369(3/31/2027), REKZ0376(5/31/2027), RELN1835(6/30/2027); 5605300/801741086984/ REKR3428(9/30/2026), REKS2486(10/31/2026), REKT3038(11/30/2026), REKV1678(10/31/2026), REKW2736(1/31/2027), REKX3721(3/31/2027), REKY2472(4/30/2027); 5615120/801741075087/ REKR0770(9/30/2026), REKR2157(9/30/2026), REKT2999(11/30/2026), REKX1764(1/31/2027), REKZ0393(2/28/2027); 5615150/801741075094/ REKQ3350(8/31/2026), REKR2193(9/30/2026), REKS1070(10/31/2026), REKS2494(10/31/2026), REKT0570(11/30/2026), REKT2132(10/31/2026), REKU1247(10/31/2026), REKU3675(12/31/2026), REKV1674(12/31/2026), REKW1228(2/28/2027), REKW2108(1/31/2027), REKX2412(3/31/2027), REKY0262(3/31/2027), REKZ0396(5/31/2027), RELN0870(6/30/2027), RELP0317(7/31/2027); 5615200/801741075100/ REKQ1130(8/31/2026), REKQ3357(8/31/2026), REKR0905(9/30/2026), REKR2215(9/30/2026), REKS1083(10/31/2026), REKT1724(11/30/2026), REKU1255(12/31/2026), REKU3680(12/31/2026), REKV2897(1/31/2027), REKW2743(2/28/2027), REKX0413(12/31/2026), REKX3745(1/31/2027), REKY2418(4/30/2027), REKZ0377(5/31/2027), RELP0308(7/31/2027); 5615240/801741075117/ REKQ3364(8/31/2026), REKR3434(9/30/2026), REKU2632(12/31/2026), REKX0415(12/31/2026), REKY0249(3/31/2027), RELN2947(6/30/2027); 5655120/801741066054/ REKV0931(1/31/2027), REKX0578(3/31/2027), REKY2420(2/28/2027); 5655150/00801741066061/ REKR2161(9/30/2026), REKV0928(10/31/2026), REKX3746(3/31/2027), REKY2398(3/31/2027), RELP1351(7/31/2027); 5655200/801741066078/ REKQ3353(8/31/2026), REKR2150(9/30/2026), REKV1673(10/31/2026), REKW2116(1/31/2027), RELN0858(6/30/2027); 5655240/801741066085/ REKS2492(10/31/2026), REKX1759(2/28/2027), REKY0238(4/30/2027); 5605200G/801741100154/ REKQ2567(8/31/2026), REKV1675(1/31/2027), REKX2411(3/31/2027), REKZ2321(5/31/2027); 5855200G/801741107146/ REKS2190(10/31/2026), REKX2413(2/28/2027), REKY0258(4/30/2027), REKZ0385(5/31/2027); 5865150G/00801741107153/ REKQ2583(8/31/2026), REKY1373(4/30/2027) 5855150/801741107412/ REKR0766(9/30/2026), REKR2162(8/31/2026), REKS2499(10/31/2026), REKU2621(12/31/2026), REKW2723(1/31/2027), REKY0252(4/30/2027), REKY2469(4/30/2027), RELN2934(6/30/2027); 5855200/801741107160/ REKQ3347(8/31/2026), REKR2211(9/30/2026), REKS3567(8/31/2026), REKU2629(12/31/2026), REKW0656(1/31/2027), REKX0406(3/31/2027), REKY0246(4/30/2027), RELP0770(7/31/2027); 5855240/801741107177/ REKQ2576(8/31/2026), REKR3432(9/30/2026), REKU3683(10/31/2026) , REKW0667(2/28/2027), REKY0240(4/30/2027), REKY2412(3/31/2027), RELN0859(6/30/2027); 5855300/801741107184/ REKQ1117(8/31/2026), REKR2204(9/30/2026), REKU1254(12/31/2026), REKY1364(4/30/2027); 5865120/801741107191/ REKS3553(10/31/2026), REKU1626(12/31/2026), REKY0241(4/30/2027), RELN0881(6/30/2027); 5865150/801741107207/ REKQ1100(8/31/2026), REKS1552(10/31/2026), REKU1594(10/31/2026), REKV2895(1/31/2027), REKX3723(3/31/2027), REKZ0397(5/31/2027); 5865200/801741107214/ REKS1549(10/31/2026), REKV0941(12/31/2026), REKY1381(4/30/2027), RELN0861(6/30/2027); 5865240/801741107221/ REKS2488(10/31/2026), REKU1253(12/31/2026), REKW2113(1/31/2027), REKY1365(4/30/2027);
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-2984-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||
| Z-2985-2026 | Class I | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.