Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3157-2026Class II

IV Non-Dedicated Gravity Administration Sets with drip chambers

ICU Medical, Inc. / initiated 2026-05-12 / Open, Classified

Recall record

Recall number
Z-3157-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
ICU Medical, Inc.
Product
IV Non-Dedicated Gravity Administration Sets with drip chambers
Product code
FMG Stopcock, I.V. Set
Reason for recall
Intravascular administration tubing set drip chambers may have black specks potentially due to the plastic molding process leaving discolored polyvinyl chloride (PVC) material within the drip chamber wall that does not enter the fluid path, which if seen by a healthcare provider may lead to replacement of product before use, which could result in a delay in therapy.
FDA root cause
Process control
Action
On 5/12/2026, the firm posted a Information Regarding Reported Black Specks in Certain IV Tubing Set Drip Chambers from ICU Medical document on their website. On 5/18/2026, updated the notices were sent. On 7/28/2026, correction notices were sent to customers asking them to do the following: 1) Share this notification with all potential users of the device to ensure they are aware of this notification. If the devices are used or further distributed to another location, please ensure that this communication is delivered there. 2) Complete and return the attached Customer Response Form to [email protected] 3) Continue using products as intended and per facility protocol. 4) Report any product quality concerns to ICU Medical Complaint Management. Firm contacts: [email protected] 1-866-261-8806. [email protected] 1-800-258-5361
Quantity in commerce
2,450
Distribution
Worldwide - US Nationwide distribution including in the states and territories of ND, TX, CT, KY, CO, AR, NC, NJ, VA, MD, OK, AK, IA, IN, KS, OR, CA, OH, WA, GA, AL, SC, MA, HI, FL, NH, NE, MN, MS, SD, MT, NY, DC, ID, WI, PA, AZ, MO, NV, TN, NM, WV, WY, IL, UT, MI, RI, VT, LA, DE, ME, VI, GU, PR and the countries of PK, PH, MY, HK, IN, DE, CR.
Product codes and lots
All products manufactured on or before 7/4/2026. UDI-DI: 10887709079363 10887709079370 10887709079417 10887709079431 10887709079448 10887709079455 10887709097473 10887709079493 10887709079509 10887709079547

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3157-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3156-2026Class IIOngoingVoluntary: Firm initiated
Z-3157-2026Class IIOngoingVoluntary: Firm initiated
Z-3158-2026Class IIOngoingVoluntary: Firm initiated
Z-3159-2026Class IIOngoingVoluntary: Firm initiated
Z-3160-2026Class IIOngoingVoluntary: Firm initiated
Z-3161-2026Class IIOngoingVoluntary: Firm initiated
Z-3162-2026Class IIOngoingVoluntary: Firm initiated