Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3253-2026Class II

nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT...

Criticare Technologies, Inc. / initiated 2026-05-06 / Open, Classified

Recall record

Recall number
Z-3253-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Criticare Technologies, Inc.
Product
nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT 8100EP1-ST, CAT 8100EP1-ST-US, CAT 8100EP1-US, CAT 8100EP1R
Product code
MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
Reason for recall
Complaint trend where "Checkline Error" alarm presents at device setup, rendering the CO2 monitoring function inoperable. May result in procedure delay.
FDA root cause
Process design
Action
Firm posted its initial notice on May 6, 2026. A follow-up letter was sent on August 31, 2026 with additional instructions. Customers were instructed to do the following: 1. Be aware of the Checkline Error alarm. If it appears on your CO2 section at startup, follow the steps provided in the letter. 2. Verify your backup plan. Confirm that your practice has access to an alternate CO2 monitoring source. 3. Retain the letter and note your unit s serial number for your records. 4. Contact the firm if you have any questions.
Quantity in commerce
1,007
Distribution
US Nationwide distribution including in the states and territory of AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NH, NJ, NV, NY, OH, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and Puerto Rico.
Product codes and lots
UDI-DI +B4388100EP1US02 All units with CO2 option SN beginning with 424, 125, 225, 325, 425, 126, 226, 326, 426, date range 2024-1st qt 2026

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3253-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3253-2026Class IIOngoingVoluntary: Firm initiated