Z-3253-2026Class II
nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT...
Criticare Technologies, Inc. / initiated 2026-05-06 / Open, Classified
Recall record
- Recall number
- Z-3253-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Criticare Technologies, Inc.
- Product
- nGenuity Patient Monitor with Capnography (ETCO2) CAT 8100E1-ST-US, CAT 8100E1-US, CAT 8100EP1, CAT 8100EP1-ST, CAT 8100EP1-ST-US, CAT 8100EP1-US, CAT 8100EP1R
- Product code
- MHX Monitor, Physiological, Patient(With Arrhythmia Detection Or Alarms)
- Reason for recall
- Complaint trend where "Checkline Error" alarm presents at device setup, rendering the CO2 monitoring function inoperable. May result in procedure delay.
- FDA root cause
- Process design
- Action
- Firm posted its initial notice on May 6, 2026. A follow-up letter was sent on August 31, 2026 with additional instructions. Customers were instructed to do the following: 1. Be aware of the Checkline Error alarm. If it appears on your CO2 section at startup, follow the steps provided in the letter. 2. Verify your backup plan. Confirm that your practice has access to an alternate CO2 monitoring source. 3. Retain the letter and note your unit s serial number for your records. 4. Contact the firm if you have any questions.
- Quantity in commerce
- 1,007
- Distribution
- US Nationwide distribution including in the states and territory of AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, ME, MI, MN, MO, MT, NC, ND, NH, NJ, NV, NY, OH, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and Puerto Rico.
- Product codes and lots
- UDI-DI +B4388100EP1US02 All units with CO2 option SN beginning with 424, 125, 225, 325, 425, 126, 226, 326, 426, date range 2024-1st qt 2026
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3253-2026 | Class II | Ongoing | Voluntary: Firm initiated |