Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators DELUXE PRE OP KIT DYKS2018 GENERAL...
Medline Industries, LP began a recall of Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators DELUXE PRE OP KIT DYKS2018 GENERAL PREOP KIT DYKS1584A PRE OP KIT DYKS1216I PRE-OP... on . The reason FDA records is "Downstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3074-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators DELUXE PRE OP KIT DYKS2018 GENERAL PREOP KIT DYKS1584A PRE OP KIT DYKS1216I PRE-OP KIT DYKM1398B
- Product code
- OJU Skin Prep Tray
- Reason for recall
- Downstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Firm began notifying consignees on 7/20/2026 via First-Class Mail and email. Medline Industries LP instructed customers to request stickers to over-label the affected convenience kits with a label indicating to remove the affected component and replace it with product from supply. No kits are to be returned.
- Quantity in commerce
- 222,653 total
- Distribution
- US Nationwide distribution.
- Product codes and lots
- DYKS2018 UDI-DI 10198459188480 Lot 25GMI043 DYKS1584A UDI-DI 10198459146633 Lots 24JBA740 25ABF244 DYKS1216I UDI-DI 10198459362668 Lots 25GBP888 25JBD789 26ABL474 26CBF334 26DBD640 26EBB855 26FBC632 DYKM1398B UDI-DI 10198459413827 Lots 25HBI183 25JBK222 25KBA990 25LBJ990 26BBI525 26CBJ249 26EBQ630 26FMF189
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3067-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3068-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3069-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3070-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3071-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3072-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3073-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3074-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3075-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3076-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3077-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3078-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3079-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.