Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators LABOR CDS CDS981762B VAG DELIVERY...
Medline Industries, LP began a recall of Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators LABOR CDS CDS981762B VAG DELIVERY ADMISSION BAG DYKA1054F on . The reason FDA records is "Downstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.
Recall record
- Recall number
- Z-3076-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Product
- Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators LABOR CDS CDS981762B VAG DELIVERY ADMISSION BAG DYKA1054F
- Product code
- MLS Labor And Delivery Kit
- Reason for recall
- Downstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.
- Root cause tag
- manufacturing process
- FDA root cause
- Nonconforming Material/Component
- Action
- Firm began notifying consignees on 7/20/2026 via First-Class Mail and email. Medline Industries LP instructed customers to request stickers to over-label the affected convenience kits with a label indicating to remove the affected component and replace it with product from supply. No kits are to be returned.
- Quantity in commerce
- 222,653 total
- Distribution
- US Nationwide distribution.
- Product codes and lots
- CDS981762B UDI-DI 10193489306767 Lots 24JBE456 24JBN932 24KBJ154 25ABK513 25CBL684 25HBK086 25IBM188 25KBL720 25LBF812 25LBU497 26ABM665 26CBH544 26CBT272 26FBK806 DYKA1054F UDI-DI 10193489312607 Lots 24ILA823 24JLA285 24LLA226 24LLB033 25BLA842 25ELA624 25GLA270 25HLB055 25KLA019 25KLA187 26BLA150 26CLA717 26DLA410 26ELA772 26FLA756
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3067-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3068-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3069-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3070-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3071-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3072-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3073-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3074-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3075-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3076-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3077-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3078-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3079-2026 | Class II | Ongoing | Voluntary: Firm initiated |