Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3077-2026Class II

Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators C-SECTION PREP DYNJ32702B

MEDLINE INDUSTRIES, LP - Northfield / initiated 2026-07-20 / Open, Classified / root cause: manufacturing process

Medline Industries, LP began a recall of Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators C-SECTION PREP DYNJ32702B on . The reason FDA records is "Downstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially...". As of , the recall status on record is Open, Classified, and MDM tags the root cause as manufacturing process.

Recall record

Recall number
Z-3077-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
MEDLINE INDUSTRIES, LP - Northfield
Product
Medline Convenience Kits containing BD ChloraPrep Clear - 1mL Applicators C-SECTION PREP DYNJ32702B
Product code
OHM Cesarean Section Tray
Reason for recall
Downstream recall - manufacturer recalled specific lots of BD ChloraPrep Clear - 1mL Applicators. Affected product may exhibit an open or incomplete seal on the packaging of the applicator. Use of potentially contaminated applicator may lead to localized skin or soft tissue infection, or systemic infection or infectious complications.
Root cause tag
manufacturing process
FDA root cause
Nonconforming Material/Component
Action
Firm began notifying consignees on 7/20/2026 via First-Class Mail and email. Medline Industries LP instructed customers to request stickers to over-label the affected convenience kits with a label indicating to remove the affected component and replace it with product from supply. No kits are to be returned.
Quantity in commerce
222,653 total
Distribution
US Nationwide distribution.
Product codes and lots
UDI-DI 10193489431070 Lots 25BBB005 25GBY063 25KBI862 25LBT688 26CBG211

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3077-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3067-2026Class IIOngoingVoluntary: Firm initiated
Z-3068-2026Class IIOngoingVoluntary: Firm initiated
Z-3069-2026Class IIOngoingVoluntary: Firm initiated
Z-3070-2026Class IIOngoingVoluntary: Firm initiated
Z-3071-2026Class IIOngoingVoluntary: Firm initiated
Z-3072-2026Class IIOngoingVoluntary: Firm initiated
Z-3073-2026Class IIOngoingVoluntary: Firm initiated
Z-3074-2026Class IIOngoingVoluntary: Firm initiated
Z-3075-2026Class IIOngoingVoluntary: Firm initiated
Z-3076-2026Class IIOngoingVoluntary: Firm initiated
Z-3077-2026Class IIOngoingVoluntary: Firm initiated
Z-3078-2026Class IIOngoingVoluntary: Firm initiated
Z-3079-2026Class IIOngoingVoluntary: Firm initiated