Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3331-2026

Brand Name: Penumbra System (RED 62 Reperfusion Catheter) Product Name: Penumbra System (RED 62 Reperfusion...

Penumbra Inc. / initiated 2026-07-24 / Open, Classified

Recall record

Recall number
Z-3331-2026
Status
Open, Classified
Initiated
Posted
Recalling firm
Penumbra Inc.
Product
Brand Name: Penumbra System (RED 62 Reperfusion Catheter) Product Name: Penumbra System (RED 62 Reperfusion Catheter) Model/Catalog Number: THNDRRED62SKIT Software Version: N/A Product Description: Product common name: Percutaneous catheter A complete product description is provided in Attachment 1 (HHE-2026-002 RED62). Component: Penumbra System
Product code
NRY Catheter, Thrombus Retriever
Reason for recall
Due to manufacturing issues reperfusion catheters may fracture.
FDA root cause
Under Investigation by firm
Action
On 07/24/2026, an "URGENT: VOLUNTARY MEDICAL DEVICE RECALL NOTIFICATION" was sent via email and FedEx Priority Overnight to customers informing them that Penumbra has identified an issue (frature) in the lot H00007856 of the Penumbra System RED 62 Reperfusion Catheter (RED 62). Actions Customers need to take: " Share this Recall Notification with all users of the product within your facility " Review inventory for the specified lot listed above " If you do have inventory of the specified lot listed above, remove affected units from inventory and place in quarantine for return to Penumbra, Inc. Complete and return the Product Identification/Return Notification to [email protected] within 3 business days. Your Penumbra Sales Representative will work with you directly to arrange return and credit. " If you do not have inventory of the specified lot listed above, complete and return the Product Identification/Return Notification form to [email protected] within 3 business days. No further action is required. For questions - contact Penumbra Sales Representative or email at [email protected].
Quantity in commerce
43
Distribution
US Nationwide distribution in the states of AL, CA, CO, FL, LA, NC, NY, TX, UT, WI.
Product codes and lots
Lot Code: UDI: 00815948024359 Lot Number: H00007856