Z-3204-2026Class II
TRIDENT 0 POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE...
Howmedica Osteonics Corp. / initiated 2026-07-30 / Open, Classified
Recall record
- Recall number
- Z-3204-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Howmedica Osteonics Corp.
- Product
- TRIDENT 0 POLY INSERT 28mm ID 620-00-28D Orthopaedic implant components manufactured from UHMWPE (polyethylene)
- Product code
- MEH Prosthesis, Hip, Semi-Constrained, Uncemented, Metal / Polymer, Non-Porous, Calcium Phosphate
- Reason for recall
- Affected devices were manufactured with Ultra High Molecular Weight Polyethylene (UHMWPE) raw material aged in excess of five years. Product has the potential for elevated levels of oxidation, which may result in increased wear, leading to hazards such as excessive wear debris, material fragments, fractured device, or disassociated device. Potential harms include revision surgery, pain, and inflammation.
- FDA root cause
- Process change control
- Action
- Firm began notifying customers on July 30, 2026, via FedEx 2-day air. Customers were instructed to quarantine and discontinue use of the affected product. Affected product should be returned to the firm.
- Quantity in commerce
- 12,156 US
- Distribution
- Worldwide distribution - US Nationwide and the countries of Canada, LATAM, Europe, Asia Pacific, and South Africa.
- Product codes and lots
- 620-00-28D UDI-DI 07613327313581 Lots MHANP2 MHD1N0
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3195-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3196-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3197-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3198-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3199-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3200-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3201-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3202-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3203-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3204-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3205-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3206-2026 | Class II | Ongoing | Voluntary: Firm initiated |