Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3...

GE Medical Systems Information Technologies Inc / initiated 2026-07-30 / Open, Classified

Recall record

Recall number
Z-3143-2026
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
GE Medical Systems Information Technologies Inc
Product
Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV
Product code
DQK Computer, Diagnostic, Programmable
Reason for recall
Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.
FDA root cause
Under Investigation by firm
Action
On July 30, 2026, the firm sent Urgent Medical Device Correction letters to customers with affected units. This letter informs the customer of the issue and provides actions to be taken by the customer/user for continued use of the device. GE HealthCare will provide a User Manual Addendum that provides guidance regarding the prolonged startup condition.
Quantity in commerce
29,570 (23,632 US; 5,938 OUS)
Distribution
Worldwide - US Nationwide distribution.
Product codes and lots
5867474-04 UDI-DI 00195278343512 Device Serial Numbers begin with STU 5867474-13 UDI-DI 00195278516367 Device Serial Numbers begin with STU 5867474-03 UDI-DI 00195278343383 Device Serial Numbers begin with STV

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3143-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3143-2026Class IIOngoingVoluntary: Firm initiated