Z-3143-2026Class II
Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3...
GE Medical Systems Information Technologies Inc / initiated 2026-07-30 / Open, Classified
Recall record
- Recall number
- Z-3143-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- GE Medical Systems Information Technologies Inc
- Product
- Carescape Central Station CSCS V3 MAS800 Desktop Part Number 5867474-04, 5867474-13 Product code STU CSCS V3 MAI800 Integrated Part Number 5867474-03 Product Code STV
- Product code
- DQK Computer, Diagnostic, Programmable
- Reason for recall
- Device may not resume patient monitoring for at least 3 minutes following system restart. Issue can lead to potential delay in recognition of clinically significant events.
- FDA root cause
- Under Investigation by firm
- Action
- On July 30, 2026, the firm sent Urgent Medical Device Correction letters to customers with affected units. This letter informs the customer of the issue and provides actions to be taken by the customer/user for continued use of the device. GE HealthCare will provide a User Manual Addendum that provides guidance regarding the prolonged startup condition.
- Quantity in commerce
- 29,570 (23,632 US; 5,938 OUS)
- Distribution
- Worldwide - US Nationwide distribution.
- Product codes and lots
- 5867474-04 UDI-DI 00195278343512 Device Serial Numbers begin with STU 5867474-13 UDI-DI 00195278516367 Device Serial Numbers begin with STU 5867474-03 UDI-DI 00195278343383 Device Serial Numbers begin with STV
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3143-2026 | Class II | Ongoing | Voluntary: Firm initiated |