Z-3108-2026Class I
Halyard IV START KIT NON-STERILE REF: KRIV26-02
AVID Medical, Inc. / initiated 2026-07-14 / Open, Classified
Recall record
- Recall number
- Z-3108-2026
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- AVID Medical, Inc.
- Product
- Halyard IV START KIT NON-STERILE REF: KRIV26-02
- Product code
- LRO General Surgery Tray
- Reason for recall
- Convenience kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which are contaminated with Aspergillus penicillioides
- FDA root cause
- Packaging process control
- Action
- On August 18, 2026, Avid Medical Inc. a subsidiary of Owens Minor issued an "Urgent Medical Device Recall Update" to affected consignees via email. The update was due to a revise recall strategy. Initial notification was sent on July 14, 2026. Avid ask consignees to take the following actions: 1. Identify, segregate, and quarantine all impacted product identified in TABLE 1. 2. Discontinue use of the BD ChloraPrep Clear - 1 mL and FREPP" Clear 1.5 mL Applicators. 3. FOR AVID KIT KRIV26-02, IV START KIT NON-STERILE, KIT LOT 1658437: Discard entire kit as it contains the contaminated component BD ChloraPrep" Clear - 1 mL, BD LOT 4032183. 4. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 5. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified of the actions to take and maintain records of effectiveness. 6. Complete the attached Recall Response Form and return by August 28, 2026. If you have any questions, contact Manager, Regulatory Affairs at 470-280-4277.
- Quantity in commerce
- 288 units
- Distribution
- US Nationwide distribution in the state of NE.
- Product codes and lots
- Lot Number: 1658437 UDI: 10809160460481
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3108-2026 | Class I | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY