Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3108-2026Class I

Halyard IV START KIT NON-STERILE REF: KRIV26-02

AVID Medical, Inc. / initiated 2026-07-14 / Open, Classified

Recall record

Recall number
Z-3108-2026
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
AVID Medical, Inc.
Product
Halyard IV START KIT NON-STERILE REF: KRIV26-02
Product code
LRO General Surgery Tray
Reason for recall
Convenience kits contain BD ChloraPrep Clear - 1 mL and FREPP Clear 1.5 mL Applicators which are contaminated with Aspergillus penicillioides
FDA root cause
Packaging process control
Action
On August 18, 2026, Avid Medical Inc. a subsidiary of Owens Minor issued an "Urgent Medical Device Recall Update" to affected consignees via email. The update was due to a revise recall strategy. Initial notification was sent on July 14, 2026. Avid ask consignees to take the following actions: 1. Identify, segregate, and quarantine all impacted product identified in TABLE 1. 2. Discontinue use of the BD ChloraPrep Clear - 1 mL and FREPP" Clear 1.5 mL Applicators. 3. FOR AVID KIT KRIV26-02, IV START KIT NON-STERILE, KIT LOT 1658437: Discard entire kit as it contains the contaminated component BD ChloraPrep" Clear - 1 mL, BD LOT 4032183. 4. Customers (end-users): If you received product from a distributor, please contact the distributor or your sales representative with any questions you may have. 5. Distributors: If any of the affected kits identified above were further distributed, please ensure all end-users are appropriately notified of the actions to take and maintain records of effectiveness. 6. Complete the attached Recall Response Form and return by August 28, 2026. If you have any questions, contact Manager, Regulatory Affairs at 470-280-4277.
Quantity in commerce
288 units
Distribution
US Nationwide distribution in the state of NE.
Product codes and lots
Lot Number: 1658437 UDI: 10809160460481

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3108-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3108-2026Class IOngoingVoluntary: Firm initiated