VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006

Steris Corporation / initiated 2026-08-31 / Open, Classified

Recall record

Recall number
Z-0070-2027
Status
Open, Classified
Initiated
Posted
Recalling firm
Steris Corporation
Product
VERIFY Biological/Integrator Challenge Pack Model/Catalog Number: LCB006
Product code
FRC Indicator, Biological Sterilization Process
Reason for recall
Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
FDA root cause
Under Investigation by firm
Action
Firm began notifying customers on August 31, 2026 via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to destroy any affected product on hand. The firm will replace any product still within its shelf life.
Quantity in commerce
266
Distribution
Worldwide - US Nationwide distribution.
Product codes and lots
UDI: 10724995220324 Lot numbers: 260810 - expiration date 08/10/2026 260810A - expiration date 08/10/2026 261106 - expiration date 11/06/2026 270116 - expiration date 01/16/2027