Z-0073-2027
VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061
Steris Corporation / initiated 2026-08-31 / Open, Classified
Recall record
- Recall number
- Z-0073-2027
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Steris Corporation
- Product
- VERIFY Dual Species Self-Contained Biological Indicator Model/Catalog Number: S3061
- Product code
- FRC Indicator, Biological Sterilization Process
- Reason for recall
- Geobacillus stearothermophilus spore population in affected products may fall below minimum shelf life specification. When used to monitor steam sterilization cycles, the potential hazard is that the biological indicator will show no growth, indicating that the steam cycle was effective, even if the steam cycle did not achieve the intended spore reduction
- FDA root cause
- Under Investigation by firm
- Action
- Firm began notifying customers on August 31, 2026 via URGENT MEDICAL DEVICE RECALL letters. Customers were instructed to destroy any affected product on hand. The firm will replace any product still within its shelf life.
- Quantity in commerce
- 2205
- Distribution
- Worldwide - US Nationwide distribution.
- Product codes and lots
- UDI: 50724995023824 Lot numbers: 260810: expiration date - 08/10/2026 260810A: expiration date - 08/10/2026 261106: expiration date - 11/06/2026 270116: expiration date - 01/16/2027 270302: expiration date - 03/02/2027