Z-0159-2027

MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETM...

Microgenics Corporation / initiated 2026-08-17 / Open, Classified

Recall record

Recall number
Z-0159-2027
Status
Open, Classified
Initiated
Posted
Recalling firm
Microgenics Corporation
Product
MAS Omni IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, REF: OPRO-101; MAS Omni IMMUNETM PRO Liquid Assayed Integrated Immunoassay Control, Sample Pack, REF: OPRO-SP
Product code
JJY Multi-Analyte Controls, All Kinds (Assayed)
Reason for recall
Vial-to-vial variability in MAS Omni"IMMUNE PRO Liquid Assayed Integrated Immunoassay Control, Level 1, Lot OPRO29061A results in low Parathyroid Hormone (PTH) recovery, which may lead to a delay in reporting patient PTH results.
FDA root cause
Process change control
Action
On 8/17/2026, recall notices were emailed to customers asking them to do the following: 1) If the variability of the material (PTH Analyte) exceeds your laboratory's specification, dispose of the material. of the material in accordance with local, regional, and national regulations. 2) Contact [email protected] for any additional support. 3) Customers using the affected product can continue to use the product as long as it meets their laboratory's specification and continue monitoring performance of PTH analyte. 4) This recall notice needs to be passed to anyone within your organization or any organization where the potentially affected lot have been transferred. 5) Retain a copy of this letter for your laboratory records. 6) Complete and return the Recall Response Form via email to [email protected] In addition, distributors were asked to do the following: - Please notify your customers of this Recall Notice using this recall notice and request they return a response to your contact information. - Firm is providing a customer response form template as a word document. Please place this letter on to your own letterhead and collect responses from your customers on that customer response form. Please note, customers should provide their response directly to your contact information. If you have any additional questions regarding this Recall Notice, contact the firm at 1-800-232-3342 (USA and Canada only) or 510-979-5000 (International), option 2, then option 3 or by email: [email protected]
Quantity in commerce
1366
Distribution
Worldwide - US Nationwide distribution in the states of UT, MI, OK, NY, TX, CA, LA, SC, OH, MO, OR and the countries of KOR, BGD, THA, ZAF, DEU, VNM, BRA, NZL.
Product codes and lots
REF/UDI-DI/Lot(Expiration): OPRO-101/00884883005327/OPRO29061A(2029-06-30); OPRO-SP/00884883005358/OPRO2906SA(2029-06-30)