Z-0153-2027
PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5
Ethicon, LLC / initiated 2026-09-08 / Open, Classified
Recall record
- Recall number
- Z-0153-2027
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Ethicon, LLC
- Product
- PERMA-HAND Silk Black Braided Sterile Nonabsorbable Surgical Suture, 1X18" TF-5, Product Code: N271H
- Product code
- GAQ Suture, Nonabsorbable, Steel, Monofilament And Multifilament, Sterile
- Reason for recall
- Silk sutures, indicated for use in cardiovascular, ophthalmic and neurological procedures, packaged with alcohol-based fluid, not part of product design, that if discovered prior to use may cause surgical delay. If used, fluid could cause reduced suture tensile strength or knot security, tissue irritation/inflammatory tissue response, and treatment failure, including wound dehiscence/bleeding.
- FDA root cause
- Process control
- Action
- On 9/8/2026, recall notices were mailed to customers who were asked to do the following: 1) Quarantine affected products. Maintain a copy of this notice with the quarantined product and keep a copy for your records. 2) Communicate the issue to relevant operating room or materials management personnel, or anyone else in your facility who needs to be informed. If any product subject to this recall has been forwarded to another facility, contact that facility to arrange return and forward them this recall notice. 3) Complete and return the Business Reply Form via email to [email protected] 4) Return affected product using the shipping label provided. In addition distributors were told the following: Based on your reported sales data, Ethicon will send notices directly to customers about the product subject to this recall. This notice will include instructions on returning the product for credit reimbursement. Distribute this notice within your organization and to any non-customer organization that received affected product from your facility. If you have additional questions regarding this communication or want to report any product complaints, please contact the firm's Resource Department at 1-877-384-4266.
- Quantity in commerce
- 1764
- Distribution
- Worldwide - US Nationwide distribution in the states of SC, PA, TX, CA, ID, LA, MO, AL, NC, UT, IL, MA, OH, AZ, TN, OK, NE, ME, IN, WI, FL, CO, WY, NH, NY, NC, MN and the country of CA.
- Product codes and lots
- UDI-DI:Box/Unit: 30705031047369/10705031047365. Lot: RLBDPJ. Expiration: SEPTEMBER 30, 2026
FIRST ARTICLE / WEEKLY