Z-0203-2027Class II

CareLink Clinic, REF: MMT-7350

Medtronic MiniMed, Inc. / initiated 2026-09-11 / Open, Classified

Recall record

Recall number
Z-0203-2027
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
Medtronic MiniMed, Inc.
Product
CareLink Clinic, REF: MMT-7350
Product code
PHV Continuous Glucose Monitor Retrospective Data Analysis Software
Reason for recall
CareLink Clinic software version 4.4A includes a software defect in the Settings Calculator feature in which the calculation logic fails to use custom values entered by healthcare professionals (HCPs) for Total Daily Dose (TDD) or Long-acting insulin when calculating the Carb Ratio, Insulin Sensitivity Factor (ISF), and Max Recommended value for Rapid-acting insulin. Instead, the calculator continues to use the system-recommended values, resulting in incorrect rapid-acting insulin parameters. If an HCP or patient inputs these incorrect setting values into the patient's MiniMed Go app, the patient could deliver higher or lower insulin doses compared to their insulin needs, potentially leading to over/under delivery of insulin with a severity ranging from Catastrophic (death from severe hypoglycemia) to Minor (self-treatable hypoglycemia/hyperglycemia).
FDA root cause
Software design
Action
On 9/11/2026, correction notices were mailed to the health care professional, and customer asking them to do the following: The Customer was asked to do the following: Immediately contact your Healthcare Provider (HCP) to confirm your insulin settings values. Complete and return the confirmation form by visiting https://info.medtronicdiabetes.com/FA1587 The health care professional was asked to do the following: If you provided diabetes treatment recommendations to patients based on custom Total Daily Dose(TDD) or long-acting insulin values entered into the CareLin Clinic Settings Calculator between: June 04, 2026, to June 08, 2026, please revisit those recommendations utilizing patients' individual CareLink reports. Continue to follow up with those patients pursuant to your standard of care. If you have further questions or need assistance, please call the firm's 24-Hour Technical Support line at 1-800-646-4633, option 1.
Quantity in commerce
1
Distribution
US Nationwide distribution in the state of KS.
Product codes and lots
UDI-DI: 0763000B00008748K. Software Versions: 4.4A Software Available: US: Between June 4th, 2026, and June 8th, 2026.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0203-2027
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0203-2027Class IIOngoingVoluntary: Firm initiated