Z-0162-2027Class III

LeadCare II Blood Lead Test Kit. Model/Catalog Number: 70-6762

Magellan Diagnostics, Inc. / initiated 2026-09-04 / Open, Classified

Recall record

Recall number
Z-0162-2027
Classification
Class III
Status
Open, Classified
Initiated
Posted
Recalling firm
Magellan Diagnostics, Inc.
Product
LeadCare II Blood Lead Test Kit. Model/Catalog Number: 70-6762.
Product code
DOF Lead, Atomic Absorption
Reason for recall
In kit sub-lot 2530M-01 the date of expiration printed on the outer labeling may indicate the wrong expiration date of July 22, 2026 while the correct expiration date is July 13, 2026. If used the system will display an error message and prevent the operator from running the test. Resulting in delayed results.
FDA root cause
Labeling Change Control
Action
Magellan notified consignees on about September 4, 2026, via telephone call indicating a written notification of the recall would follow. On 09/17/2026, a notification letter was emailed to consignees with the following instructions: 1. Complete, sign, and return the CONFIRMATION OF NOTIFICATION FORM. 2. Review your inventory for the affected LeadCare II Blood Lead Test Kit lot identified. 3. Immediately discontinue use of any remaining affected kits. 4. Return any recalled product still in your possession to Magellan Diagnostics, Inc. for destruction.
Quantity in commerce
321 units
Distribution
US Nationwide distribution in the states of CA, IN, PA, TX, NE, IL, MA, MO, KY.
Product codes and lots
Model/Catalog Number: 70-6762. UD-DI: 00850355006000. Lot Number. 2530M-01.

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0162-2027
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0162-2027Class IIIOngoingVoluntary: Firm initiated