Z-3242-2026Class II
MR 7700 and Upgrade to MR 7700; Model Number (REF): 1. 782120, 2. 782153, 3. 782130 (Upgrade)
Philips North America Llc / initiated 2026-08-21 / Open, Classified
Recall record
- Recall number
- Z-3242-2026
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips North America Llc
- Product
- MR 7700 and Upgrade to MR 7700; Model Number (REF): 1. 782120, 2. 782153, 3. 782130 (Upgrade);
- Product code
- LNH System, Nuclear Magnetic Resonance Imaging
- Reason for recall
- A software issue will terminate an examination during the image reconstruction process if the scan parameters are manually adjusted such that reconstruction requirements exceed the available system memory.
- FDA root cause
- Software design
- Action
- On August 24, 2026 URGENT Medical Device Correction letters were sent to customers. Actions to be taken: " To prevent this issue from occurring, o Use the standard 3D Vane scan protocol from the Philips Examcard Database, or o For custom 3D Vane scan protocols, use a maximum of 480 for the reconstruction matrix and use a maximum of 167 for the number of slices. " You may continue to use affected systems in accordance with their intended use. " Circulate this notice to all users or new owners of this device so that they are aware of the issues and associated hazard/harm. " Please retain this Urgent Medical Device letter with your system(s) until the software upgrade is installed; ensure the notice is in a place likely to be seen/viewed. Actions planned by Philips to correct the problem A Philips representative will contact you to schedule time for a Field Service Engineer (FSE) to install a software upgrade to resolve the issue (reference 2026-PD-MR-006).
- Quantity in commerce
- 37
- Distribution
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Andorra, Angola, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Costa Rica, C¿te D'Ivoire, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Equatorial Guinea, Estonia, Fiji, Finland, France, French Polynesia, Georgia, Germany, Ghana, Greece, Guadeloupe, Guernsey, Honduras, Hong Kong, Hungary, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Japan, Jersey, Jordan, Kazakhstan, Kenya, Korea, Republic of, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macao, Macedonia, Madagascar, Malaysia, Maldives, Malta, Martinique, Mauritius, Mayotte, Mexico, Moldova, Monaco, Mongolia, Montenegro, Morocco, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Papua New Guinea, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sweden, Switzerland, Syrian Arab Republic, Taiwan, Tanzania, Thailand, Trinidad and Tobago, Tunisia, Turkey, Turkmenistan, Ukraine, United Arab Emirates, United Kingdom, Uzbekistan, Venezuela, Viet Nam, Yemen, Zambia;
- Product codes and lots
- 1. Model Number (REF): 782120; UDI-DI: 884838104112; Serial Numbers: 65002, 65003, 65014, 65015, 65050, 65029, 65056, 65018, 65021, 65035, 65038, 65000, 65025, 65006, 65008, 65043; 2. Model Number (REF): 782153; UDI-DI: 884838112858; Serial Numbers: 68011, 68022, 68031, 68014, 68012, 68009, 68033, 68053, 68023; 3. Model Number (REF): 782130; UDI-DI: 884838104402; Serial Numbers: 45007, 45052, 45043, 45044, 45552, 71616, 42164, 42066, 42104, 42248, 45263, 71532;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3232-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3233-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3234-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3235-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3236-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3237-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3238-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3239-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3240-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3241-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3242-2026 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-3243-2026 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.