Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A

Olympus Corporation of the Americas / initiated 2026-08-24 / Open, Classified

Recall record

Recall number
Z-3299-2026
Status
Open, Classified
Initiated
Posted
Recalling firm
Olympus Corporation of the Americas
Product
Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.
Product code
HIH Hysteroscope (And Accessories)
Reason for recall
There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.
FDA root cause
Process control
Action
Olympus notified the single Japanese distributor on about 08/24/2026 via emailed "URGENT: FIELD SAFETY NOTICE." Consignees and end users are being instructed to store a copy of the provided correct sterilization parameters with current package insert, use only the corrected sterilization parameters, complete and return the provided reply form, and if units have been further distributed, notify those customers.
Quantity in commerce
72 units
Distribution
International distribution to the country of Japan.
Product codes and lots
Model Number: M3-30A. UDI-DI: 00821925005679. All Serial Numbers.