Z-3299-2026
Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A
Olympus Corporation of the Americas / initiated 2026-08-24 / Open, Classified
Recall record
- Recall number
- Z-3299-2026
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Olympus Corporation of the Americas
- Product
- Olympus M3 Gold Autoclavable Foroblique Telescope 30. Model Number: M3-30A.
- Product code
- HIH Hysteroscope (And Accessories)
- Reason for recall
- There are identified discrepancies between the sterilization parameters listed in the Instructions for Use and the sterilization parameters listed in the Japanese package insert.
- FDA root cause
- Process control
- Action
- Olympus notified the single Japanese distributor on about 08/24/2026 via emailed "URGENT: FIELD SAFETY NOTICE." Consignees and end users are being instructed to store a copy of the provided correct sterilization parameters with current package insert, use only the corrected sterilization parameters, complete and return the provided reply form, and if units have been further distributed, notify those customers.
- Quantity in commerce
- 72 units
- Distribution
- International distribution to the country of Japan.
- Product codes and lots
- Model Number: M3-30A. UDI-DI: 00821925005679. All Serial Numbers.