Z-0082-2027Class II
Sperm Ab ELISA; Model/Catalog Number: EIA-1826; Product Description: Sperm Ab ELISA (96 well plate test)
DRG International, Inc. / initiated 2026-07-02 / Open, Classified
Recall record
- Recall number
- Z-0082-2027
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- DRG International, Inc.
- Product
- Sperm Ab ELISA; Model/Catalog Number: EIA-1826; Product Description: Sperm Ab ELISA (96 well plate test);
- Product code
- DFQ Sperm, Antigen, Antiserum, Control
- Reason for recall
- Potential for false negative patient results.
- FDA root cause
- Reprocessing Controls
- Action
- On July 2, 2026 Field Safety Notice letters were sent to customers. Actions to be taken: Please check your stocks for the lot in question (320K046), stop using these kits immediately and quarantine these products without delay. Please destroy the kits and send us the confirmation with the completed response form. If you have transferred any of the affected products to another laboratory, please provide them with a copy of this letter.
- Quantity in commerce
- 211
- Distribution
- Worldwide - US Nationwide distribution in the state of NJ and the countries of Azerbaijan, Belgium, Czech Republic, Germany, Hungary, India, Italy, Kuwait, Russia, Spain, Ukraine.
- Product codes and lots
- Model/Catalog Number: EIA-1826; UDI: 04048474018264; Lot# 320K046;
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0082-2027 | Class II | Ongoing | Voluntary: Firm initiated |