Intellijoint Navigation System: Workstation G2 Configured, REF: 000-0241; Intellijoint HIP Cart System, REF...

Intellijoint Surgical, Inc. / initiated 2026-08-28 / Open, Classified

Recall record

Recall number
Z-0119-2027
Status
Open, Classified
Initiated
Posted
Recalling firm
Intellijoint Surgical, Inc.
Product
Intellijoint Navigation System: Workstation G2 Configured, REF: 000-0241; Intellijoint HIP Cart System, REF: 000-0250; Intellijoint HIP, Portable System, REF: 000-0249
Product code
OLO Orthopedic Stereotaxic Instrument
Reason for recall
A navigation system used to aid in selection and positioning of orthopedic implant system components has a workstation that may contain a 15-inch MacBook Pro battery, subject to an Apple battery recall, that may overheat and pose a fire safety risk.
FDA root cause
Nonconforming Material/Component
Action
On 8/28/2026, recall notices were emailed to customers who were asked to do the following: - If affected serial numbers are found, stop use and contact firm's customer service: at [email protected] and copy [email protected] - If the observed serial number does not match the affected serial number please record the serial number identified and inform [email protected] and cc [email protected] of the correct serial number for that facility.
Quantity in commerce
8
Distribution
Worldwide - US Nationwide distribution in the states of CA, TX, MS, NJ and the countries of Singapore, Spain.
Product codes and lots
REF/UDI-DI/MacBook Serial Numbers: 000-0241/00628184002414/C02RP36YG8WP, C02QPCF5G8WN 000-0250/00628184002506/C02RF44TG8WN, C02QW25RG8WL, C02QH355G8WM, C02S22N9G8WP; 000-0249/00628184002490/C02RQ3FUG8WN, C02R4451G8WN