Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/...
Medline Industries, LP / initiated 2026-08-14 / Open, Classified
Recall record
- Recall number
- Z-3227-2026
- Classification
- Class I
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medline Industries, LP
- Product
- Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/ MDR107002E MDR520094 PENDANT F/ MDR107003E MDR530094 PENDANT, W/ 9V BATTERY BACKUP, F/ 2E BED MDR540094 PENDANT, W/ 9V BATTERY BACKUP, F/ 3E BED
- Product code
- LLI Bed, Therapeutic, Ac-Powered, Adjustable Home-Use
- Reason for recall
- Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.
- FDA root cause
- Device Design
- Action
- Firm initially notified customers November 26, 2025. On August 14, 2026, the firm sent a follow-up letter which provides additional product SKU - Item numbers for the electrical system safety risks and additional important information related to the previous notices. The notice explained the issue, potential risks, and requested the user ONLY USE MEDLINE SIDE RAILS and MEDLINE ACCESSORIES that are intended for use with Medline Homecare Beds. When using Medline Side Rails and Medline Accessories, follow all applicable instructions for use. Distributors were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 [email protected].
- Quantity in commerce
- 771,127 total
- Distribution
- Distribution US Nationwide including territories of Puerto Rico and Guam and the countries of Bermuda, Canada, and Panama.
- Product codes and lots
- MDR510094 UDI-DI 10888277363021 MDR520094 UDI-DI 10884389161692 MDR530094 UDI-DI 10197344119059 MDR540094 UDI-DI 10197344119066 All lots
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-3227-2026 | Class I | Ongoing | Voluntary: Firm initiated | ||