Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3227-2026Class I

Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/...

Medline Industries, LP / initiated 2026-08-14 / Open, Classified

Recall record

Recall number
Z-3227-2026
Classification
Class I
Status
Open, Classified
Initiated
Posted
Recalling firm
Medline Industries, LP
Product
Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories, Hand Pendants: MDR510094 PENDANT F/ MDR107002E MDR520094 PENDANT F/ MDR107003E MDR530094 PENDANT, W/ 9V BATTERY BACKUP, F/ 2E BED MDR540094 PENDANT, W/ 9V BATTERY BACKUP, F/ 3E BED
Product code
LLI Bed, Therapeutic, Ac-Powered, Adjustable Home-Use
Reason for recall
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.
FDA root cause
Device Design
Action
Firm initially notified customers November 26, 2025. On August 14, 2026, the firm sent a follow-up letter which provides additional product SKU - Item numbers for the electrical system safety risks and additional important information related to the previous notices. The notice explained the issue, potential risks, and requested the user ONLY USE MEDLINE SIDE RAILS and MEDLINE ACCESSORIES that are intended for use with Medline Homecare Beds. When using Medline Side Rails and Medline Accessories, follow all applicable instructions for use. Distributors were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 [email protected].
Quantity in commerce
771,127 total
Distribution
Distribution US Nationwide including territories of Puerto Rico and Guam and the countries of Bermuda, Canada, and Panama.
Product codes and lots
MDR510094 UDI-DI 10888277363021 MDR520094 UDI-DI 10884389161692 MDR530094 UDI-DI 10197344119059 MDR540094 UDI-DI 10197344119066 All lots

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-3227-2026
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-3227-2026Class IOngoingVoluntary: Firm initiated
Z-3228-2026Class IOngoingVoluntary: Firm initiated