Medical Device
Manufacturing
THE TRADE PAPER OF THE FACTORY FLOOR
Z-3228-2026

Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED, MANUAL, MEDLITE

Medline Industries, LP / initiated 2026-08-14 / Open, Classified

Recall record

Recall number
Z-3228-2026
Status
Open, Classified
Initiated
Posted
Recalling firm
Medline Industries, LP
Product
Medline Basic Homecare Beds and Medlite Homecare Beds and Accessories MDR107000P BED, MANUAL, MEDLITE
Product code
FNJ Bed, Manual
Reason for recall
Potential fire hazard involving the hand control pendant and associated wiring used with affected beds. Pendant or pendant wiring may overheat, potentially leading to smoking, sparking, melting, burning, or fire. Additional products added to 11/26/2025 recall.
FDA root cause
Device Design
Action
Firm initially notified customers November 26, 2025. On August 14, 2026, the firm sent a follow-up letter which provides additional product SKU - Item numbers for the electrical system safety risks and additional important information related to the previous notices. The notice explained the issue, potential risks, and requested the user ONLY USE MEDLINE SIDE RAILS and MEDLINE ACCESSORIES that are intended for use with Medline Homecare Beds. When using Medline Side Rails and Medline Accessories, follow all applicable instructions for use. Distributors were directed to notify their customers. For questions, contact the Recall Department at 866-359-1704 [email protected].
Quantity in commerce
771,127 total
Distribution
Distribution US Nationwide including territories of Puerto Rico and Guam and the countries of Bermuda, Canada, and Panama.
Product codes and lots
MDR107000P UDI-DI 00080196288284 All lots