Z-0128-2027

VCSN+Pro Model Number: PRT-01184-000, Package PN:PRT-01172-000. used for ventilatory support

Ventec Life Systems, Inc. / initiated 2026-09-17 / Open, Classified

Recall record

Recall number
Z-0128-2027
Status
Open, Classified
Initiated
Posted
Recalling firm
Ventec Life Systems, Inc.
Product
VCSN+Pro Model Number: PRT-01184-000, Package PN:PRT-01172-000. used for ventilatory support.
Product code
CBK Ventilator, Continuous, Facility Use
Reason for recall
Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
FDA root cause
Under Investigation by firm
Action
On September 17, 2026, Ventec Life Systems Inc. dba React Health issued a Urgent: Medical Device Recall Notification via certified mail. Ventec ask consignees to take the following actions: 1.Review the customer communication and follow the applicable Actions to be Taken section. 2. Confirm receipt of this notification letter by returning the enclosed Customer Response Form to React Health via email, mail or use the online customer response form within 7 days. 3. If you have distributed or transferred affected devices to other facilities, caregivers, or end users, you must promptly notify them and provide a copy of this notification.
Quantity in commerce
4,368 units
Distribution
Distribution US nationwide including Puerto Rico.
Product codes and lots
All Serial Numbers: UDI: 00855573007860