Z-0131-2027
VC(Japan) Model Number: PRT-01098-100, Package PN:PRT-01125-100. used for ventilatory support
Ventec Life Systems, Inc. / initiated 2026-09-17 / Open, Classified
Recall record
- Recall number
- Z-0131-2027
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Ventec Life Systems, Inc.
- Product
- VC(Japan) Model Number: PRT-01098-100, Package PN:PRT-01125-100. used for ventilatory support.
- Product code
- CBK Ventilator, Continuous, Facility Use
- Reason for recall
- Ventilator's Internal Flow transducer cable connector may disconnect from internal board, potentially resulting in system reboot and interruption of ventilation therapy.
- FDA root cause
- Under Investigation by firm
- Action
- On September 17, 2026, Ventec Life Systems Inc. dba React Health issued a Urgent: Medical Device Recall Notification via certified mail. Ventec ask consignees to take the following actions: 1.Review the customer communication and follow the applicable Actions to be Taken section. 2. Confirm receipt of this notification letter by returning the enclosed Customer Response Form to React Health via email, mail or use the online customer response form within 7 days. 3. If you have distributed or transferred affected devices to other facilities, caregivers, or end users, you must promptly notify them and provide a copy of this notification.
- Quantity in commerce
- 180 units
- Distribution
- Distribution US nationwide including Puerto Rico.
- Product codes and lots
- All Serial Numbers: UDI: 00850018761116