HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)

Instrumentation Laboratory / initiated 2026-09-03 / Open, Classified

Recall record

Recall number
Z-0002-2027
Status
Open, Classified
Initiated
Posted
Recalling firm
Instrumentation Laboratory
Product
HemoslL AcuStar HIT-lgG(PF-4). Part Number: 0009802028. in vitro diagnostic (OUS only)
Product code
LCO Platelet Factor 4 Radioimmunoassay
Reason for recall
Potential for invalid calibration curve or negative mean bias of up to 47% with patient samples.
FDA root cause
Under Investigation by firm
Action
Instrumentation Laboratory notified consignees on about 09/03/2026 via emailed and mailed URGENT MEDICAL DEVICE REALL letter. Consignees were made aware of the issue and hazard involved. They were instructed to check inventory for affected lots, destroy any remaining product following each facility's standard procedures, and document the destruction on the enclosed Urgent Mandatory Response Tracking Form, including quantity destroyed and date of destruction. Consignees were asked to identify an alternative method for HIT antibody testing as a supply interruption is expected and work local Werfen representatives for guidance and transition support. Consignees were also requested to inform all laboratory personnel of the notification, forward the notification to all affected departments and locations within each facility, retain a copy of the notification for records and complete and return the Urgent Mandatory Response Tracking Form within 10 days of receipt even if there is no remaining inventory. On 09/29/2026, an expanded URGENT MEDICAL DEVICE RECALL Notification was sent to consignees through registered mail or email. The updated notification informed consignees of additional lot numbers affected. Instructions remained the same.
Quantity in commerce
1,414 units
Distribution
Worldwide - US Nationwide distribution in the states of AR, AZ, CA, CT, FL, GA, MA, MD, NJ, NY, PA, RI, TN, WA, WV and the countries of Austria, Belgium, Switzerland, Costa Rica, Czech Republic, Germany, Denmark, Spain, Finland, France, United Kingdom, Greece, India, Italy, Jordan, Japan, South Korea, Luxembourg, Latvia, Netherlands, Saudi Arabia, Sweden, Singapore.
Product codes and lots
Part Number: 0009802028. UDI-DI: 08426950545774. Lot Number (Expiration Date): B40006 (01/16/2027), B38063 (11/01/2026).