Z-0191-2027Class II
Amplatzer Amulet Left Atrial Appendage Occluder Model Numbers 1. 9-ACP2-007-016 2. 9-ACP2-007-018 3...
Abbott / initiated 2026-09-10 / Open, Classified
Recall record
- Recall number
- Z-0191-2027
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Abbott
- Product
- Amplatzer Amulet Left Atrial Appendage Occluder Model Numbers 1. 9-ACP2-007-016 2. 9-ACP2-007-018 3. 9-ACP2-007-020 4. 9-ACP2-007-022 5. 9-ACP2-010-025 6. 9-ACP2-010-028 7. 9-ACP2-010-031 8. 9-ACP2-010-034
- Product code
- NGV System, Appendage Closure, Left Atrial
- Reason for recall
- Changes to Instructions for Use (IFU) for the device regarding risk of Pulmonary Artery Injury (PAI).
- FDA root cause
- Labeling design
- Action
- US IFU was updated and is available via https://manuals.eifu.abbott as of September 4, 2026. The firm began notifying consignees of the updated IFU on September 10, 2026. Abbott asks that physician users continue to follow all procedural steps that have already been provided via training, and that are now included in the IFU, as presented below: " After the landing zone has been identified, assess the closest distance to the pulmonary artery. " A distance =1.5 mm represents significant risk for a pulmonary artery injury; a distance between 1.5 and 3.0 mm represents a theoretical risk for a pulmonary artery injury; and a distance >3.0 mm represents a lower risk for a pulmonary artery injury. " To reduce the risk of pulmonary artery injury, consider an alternative landing zone or device orientation when clinically appropriate and C.L.O.S.E. criteria can still be met. " If the risk of pulmonary artery injury cannot be adequately minimized, the physician should carefully consider whether to proceed with device implantation.
- Distribution
- Worldwide - US Nationwide distribution.
- Product codes and lots
- UDI-DI 1. 00811806013459 2. 00811806013466 3. 00811806013473 4. 00811806013480 5. 00811806013497 6. 00811806013503 7. 00811806013510 8. 00811806013527 All lots/serial numbers.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0191-2027 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY