Z-0314-2027
Precision Pointer Electromagnetic (8000-050-001)
Stryker Leibinger GmbH & Co. KG / initiated 2026-09-24 / Open, Classified
Recall record
- Recall number
- Z-0314-2027
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Stryker Leibinger GmbH & Co. KG
- Product
- Precision Pointer Electromagnetic (8000-050-001)
- Product code
- HAW Neurological Stereotaxic Instrument
- Reason for recall
- Device may fracture during use at the laser welded joint on the pointer tube. If this happens, loose parts may fall out of the product while the device is inside the patient.
- FDA root cause
- Process control
- Action
- The firm began notifying customers on September 24, 2026 via Urgent Medical Device Recall letters. Customers were instructed to immediately locate and quarantine any affected products at their facility. Customers may either destroy affected products onsite with their Stryker representative and send a completed Certificate of Destruction to the firm, or return the products to Stryker. The firm will provide a replacement upon receipt of the recalled product or Certificate of Destruction.
- Quantity in commerce
- 513
- Distribution
- Worldwide - US Nationwide distribution in the states of Alabama, Arizona, California, Colorado, Connecticut, Delaware, Florida, Hawaii, Illinois, Indiana, Kentucky, Louisiana, Massachusetts, Michigan, Missouri, Montana, New Hampshire, New York, Ohio, Pennsylvania, South Carolina, South Dakota, Tennessee, Texas, Utah, Washington and the countries of Austria, Canada, Denmark, Finland, France, Germany, Ireland, Italy, Japan, Norway, Portugal, South Korea, Spain, Switzerland, UK.
- Product codes and lots
- Lot 7000413361 UDI-DI 07613327413205