Z-0317-2027
Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test...
Accriva Diagnostics, Inc. / initiated 2026-09-14 / Open, Classified
Recall record
- Recall number
- Z-0317-2027
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Accriva Diagnostics, Inc.
- Product
- Hemochron Jr., Whole Blood Microcoagulation System, Activated Partial Thromboplastin Time (APTT) Test Cuvette, REF: J103
- Product code
- JPA System, Multipurpose For In Vitro Coagulation Studies
- Reason for recall
- Microcoagulation test used to evaluate low dose heparin anticoagulation has a mispositioned or absent clot activating reagent, which may result in out-of-range high error code, or erroneously high results, which may lead retest, or overestimation of patient's anticoagulation, leading to inadequate anticoagulant therapy, which could result in thrombosis.
- FDA root cause
- Process control
- Action
- On 9/14/2026, recall notices were mailed to customers who were informed of the following: 1) Please forward this communication to those within your organization who need to be aware of this matter and verify they understand the procedure and can complete the enclosed forms. 2) Forward the enclosed customer notification to all other organizations where affected devices may have been transferred to. 3) Please check your inventory and do not use affected lots. 4) Scrap/destroy all impacted cuvettes and/or boxes. 5) Complete and return the Acknowledgement and Receipt Form via email to [email protected] Questions about receiving credit, should be directed to Customer Service at 1-800-955-9525 or e-mail [email protected]. Technical questions should be directed to Technical Services at 1-800-678-0710 Option 3 or email [email protected].
- Quantity in commerce
- 28710
- Distribution
- Worldwide - US Nationwide distribution in the states of MA, MO, AK, WV, NJ, TX, VA, IL, NY and the countries of China, Colombia.
- Product codes and lots
- UDI-DI: 10711234103012, Lot(Expiration): B6JCA005(02/28/2027), C6JCA008(03/31/2027), D6JCA011(04/30/2027), K5JCA041(10/31/2026), C6JCA007(03/31/2027)
FIRST ARTICLE / WEEKLY
What cleared, what was recalled, and what changed in the standards. One edition a week for the people who make medical devices.
