Z-0146-2027
iFuse TORQ Implant System Model/Catalog Number: 10055T(10x55mm), 10060T(10x60mm), 10065T(10x65mm)...
SI-BONE, Inc. / initiated 2026-08-21 / Open, Classified
Recall record
- Recall number
- Z-0146-2027
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- SI-BONE, Inc.
- Product
- iFuse TORQ Implant System Model/Catalog Number: 10055T(10x55mm), 10060T(10x60mm), 10065T(10x65mm), 10080T(10x55mm), 11580T(11.5x80mm)
- Product code
- OUR Sacroiliac Joint Fixation
- Reason for recall
- The pelvic fixation screws failed to meet acceptance criteria for tube packaging seal integrity testing; therefore, sterility assurance cannot be ensured.
- FDA root cause
- Packaging
- Action
- On August 31, 2026 and September 2, 2026, SI-Bone, Inc. Chief Medical Officer spoke with affected surgeons via phone concerning the recall.
- Quantity in commerce
- 100 units
- Distribution
- US Nationwide distribution in the states of FL, GA, HI, NC, OH, PA, SD, TX, VA.
- Product codes and lots
- 10055T-UDI:00810055520077/Lot:9142871 10060T-UDI:00810055520084/Lot:9137761 10065T-UDI:00810055520091/Lot:9139213 10080T-UDI:00810055520121/Lot:9137781 11580T-UDI:00810055520367/Lot:9142911