Z-0192-2027
Think Surgical Square Driver REF: 109126
Think Surgical, Inc. / initiated 2026-08-20 / Open, Classified
Recall record
- Recall number
- Z-0192-2027
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Think Surgical, Inc.
- Product
- Think Surgical Square Driver REF: 109126
- Product code
- OLO Orthopedic Stereotaxic Instrument
- Reason for recall
- Robotic system square drive failed to meet design specifications potentially resulting in pin becoming stuck in square driver or the pin wobbling during drilling.
- FDA root cause
- Process control
- Action
- On August 25, 2026, THINK Surgical Inc. issued a Urgent Medical Device Recall Notification to affected consignees via email. Think Surgical sent an updated letter via email on September 22, 2026 to provide additional product information. THINK Surgical ask consignees to take the following actions: 1. Examine your inventory of square drivers for the lot number 25356. Quarantined return any inventory of that lot to THINK Surgical, Inc., 2. If you have distributed or transferred affected devices to other facilities or end users, please promptly notify them and provide a copy of this notification., 3. Fill out the accompanying Acknowledgement Form. Email the completed Acknowledgement Form to THINK Surgical, Inc at [email protected].
- Quantity in commerce
- 15 units
- Distribution
- US Nationwide distribution in the states of NY and DE.
- Product codes and lots
- Lot: 25356/Primary DI: M9461091260