Z-0192-2027

Think Surgical Square Driver REF: 109126

Think Surgical, Inc. / initiated 2026-08-20 / Open, Classified

Recall record

Recall number
Z-0192-2027
Status
Open, Classified
Initiated
Posted
Recalling firm
Think Surgical, Inc.
Product
Think Surgical Square Driver REF: 109126
Product code
OLO Orthopedic Stereotaxic Instrument
Reason for recall
Robotic system square drive failed to meet design specifications potentially resulting in pin becoming stuck in square driver or the pin wobbling during drilling.
FDA root cause
Process control
Action
On August 25, 2026, THINK Surgical Inc. issued a Urgent Medical Device Recall Notification to affected consignees via email. Think Surgical sent an updated letter via email on September 22, 2026 to provide additional product information. THINK Surgical ask consignees to take the following actions: 1. Examine your inventory of square drivers for the lot number 25356. Quarantined return any inventory of that lot to THINK Surgical, Inc., 2. If you have distributed or transferred affected devices to other facilities or end users, please promptly notify them and provide a copy of this notification., 3. Fill out the accompanying Acknowledgement Form. Email the completed Acknowledgement Form to THINK Surgical, Inc at [email protected].
Quantity in commerce
15 units
Distribution
US Nationwide distribution in the states of NY and DE.
Product codes and lots
Lot: 25356/Primary DI: M9461091260