Z-0152-2027
Philips Lumify Diagnostic Ultrasound System
Philips Ultrasound, LLC / initiated 2026-09-24 / Open, Classified
Recall record
- Recall number
- Z-0152-2027
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Philips Ultrasound, LLC
- Product
- Philips Lumify Diagnostic Ultrasound System
- Product code
- IYN System, Imaging, Pulsed Doppler, Ultrasonic
- Reason for recall
- Ultrasound system may: 1) Not recognize previous transducer registration, prevent scanning, require re-registration with Wi-Fi/cellular access, which may be unavailable outside of the hospital (ambulances/aircraft/field); 2) At connection between transducer USB cable to tablet/phone, there may be an unstable connection/may disconnect disrupting function. May cause diagnosis/treatment delay.
- FDA root cause
- Device Design
- Action
- On 9/24/2026, correction notices were mailed to customers who were asked to do the following: Until software update is available and installed: " Do not use the affected systems outside of areas with readily accessible Wi-Fi/Cellular (e.g. prehospital transport (ambulance, helicopter) or use outside of a traditional healthcare setting (care in the field). " Always have backup ultrasound devices and/or modalities available for care as advised in the User Guide. " Circulate this notice to all users of the device so that they are aware of the product issue and associated hazard / harm. " Post this notice near the affected unit(s) for ease of reference. " Forward this notice to any area/facility that may have the device. Additionally: 1) Complete and return the response form via email to [email protected] 2) Firm is developing a software update to address the identified registration issue. Firm is also assessing the unstable cable connection issue to identify the appropriate solution. Firm will communicate once the software update is available and provide instructions on any actions required. Questions can be directed to the firm's local representatives or to the Customer Care Solutions Center (1-800-722-9377).
- Quantity in commerce
- 13,206
- Distribution
- Worldwide - US Nationwide distribution including in the states of CA, MA, IL, NY, OR, GA, TX, WA, FL, PA, WI, LA, NJ, KY, VA, MS, AZ, SC, CT, MN, UT, OH, AL, MI, NH, MO, TN, IN, DC, NE, NM, MT, NV, NC, MD, IA, SD, VT, KS, AK, CO, ID, OK, WV, AR, RI, ME, WY, HI, DE, ND and the countries of CA, NO, HR, DE, KR, GB, ES, FI, PL, AU, TH, SG, NL, RO, CH, FR, MM, MX, BE, DK, MY, SE, LB, SI, BR, CL, GR, PH, CN, ID, LT, HK, MQ, JP, IT, IE, IL, AT, TW, AE, SA, HU, RU, VN, NZ, TR, SK, ZA, KE, AR, NP, PK, UY, DO, MN, LK, GT, BD, IS, IN, NI, LY, CY, CZ, SV, PA, BG, PE, EC, CO, VE, UZ, YE, KW, DZ, LU, EG, EE, UA, LV, KZ, TZ, MD, JO, GY, QA, MU, AM, OM, PT, TN, NG, BO, PF, ZM, GF, YT, ME, GD, RS, BW, GP, RE, MT, CR, MA, CI, NC, BA, SR, CG, TL, GH, AL, MV, XK, BN, LS, MK, CM, TT, ZW, BH, PS, HN, MC, WS, JE, PM, AZ.
- Product codes and lots
- UDI-DI/Software Version: Android: 00884838122116/1.X, 3.X, 4.X, 5.X. iOS: 00884838124141/2.X and 5.X