Z-0166-2027Class II
Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514
Howmedica Osteonics Corp. / initiated 2026-09-16 / Open, Classified
Recall record
- Recall number
- Z-0166-2027
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Howmedica Osteonics Corp.
- Product
- Triathlon Hinge Femoral Prep Tray. Catalog Number: 6543-6-514.
- Product code
- FSM Tray, Surgical, Instrument
- Reason for recall
- Reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes for the trays sterilized in blue wrap. When blue wrap is used, the 30-minute dry time may lead to a small amount of residual moisture remaining after sterilization.
- FDA root cause
- Labeling design
- Action
- Stryker notified consignees on about 09/16/2026 via "MEDICAL DEVICE CORRECTION" letter mailed through UPS 2-day air. Consignees were informed that reprocessing instructions provided in LSTPI-B specify a sterilization dry time of 30 minutes, however, these trays require a sterilization dry time of 40 minutes when sterilized in blue wrap. Consignees were instructed to inform users of the Medical Device Correction and forward the notice to all individuals who need to be made aware or organizations who have consigned product, review and follow the updated instructions provided in LSTPI-B Cleaning, Sterilization, Inspection, and Maintenance of Reusable Medical Devices Reference Guide, complete and return the provided Medical Device Correction Business Reply Form.
- Quantity in commerce
- 359 trays
- Distribution
- These trays are provided by sale or loan. Loans are provided through Distributor. Foreign Consignees will be identified by the countries impacted. 158 trays have been distributed OUS.
- Product codes and lots
- Catalog Number: 6543-6-514. GTIN: 07613327562859. All lot numbers used prior to the correction date of May 8, 2026 are impacted.
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0164-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0165-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0166-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0167-2027 | Class II | Ongoing | Voluntary: Firm initiated |