Z-0211-2027Class II
Ion Torrent Genexus Dx with software, Catalog: A40003539, which supports the Ion Torrent Genexus Dx...
Life Technologies Holdings Pte Ltd / initiated 2026-08-14 / Open, Classified
Recall record
- Recall number
- Z-0211-2027
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Life Technologies Holdings Pte Ltd
- Product
- Ion Torrent Genexus Dx with software, Catalog: A40003539, which supports the Ion Torrent Genexus Dx Integrated Sequencer
- Product code
- PFF High Throughput Dna Sequence Analyzer
- Reason for recall
- DNA or RNA sequencing system software has vulnerabilities.
- FDA root cause
- Software design
- Action
- On 8/14/2026, recall notices were emailed to customers who were informed of the following: Once the software becomes available, your field support specialist or your local Service and Support representative will reach out to schedule a software upgrade. Firm recommends that customers back up their data before any software upgrades. Please refer to "Manage backup settings and restore runs (administrator)" section of the user guide. Complete and return the CUSTOMER RESPONSE Acknowledgement and Receipt Form via email to [email protected]; and also contact the firm at this email address is you have any technical questions or concerns.
- Quantity in commerce
- 16
- Distribution
- US Nationwide distribution in the states of PA, NC, CO, CA, LA, KY, NY, AZ, MA, LA, OK.
- Product codes and lots
- UDI-DI: 10190302018883, Software: 6.8.2.1 and earlier. Sequencer UDI-DI: 10190302017794. Sequencer Serial Numbers: 2619124040913, 2619726010247, 2619124060941, 2619124060942, 2619725090229, 2619725100238, 2619122070666, 2619725080220, 2619725090226, 2619726010251, 2619726020265, 2619725090225, 2619725100235, 2619725050205, 2619726010249, 2619122070658
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0211-2027 | Class II | Ongoing | Voluntary: Firm initiated |
FIRST ARTICLE / WEEKLY