Z-0175-2027
Medline surgical convenience kits labeled as: 1) CV BASIN PACK-LF, Medline Product number/SKU DYNJ0538615K...
Medline Industries, LP / initiated 2026-09-02 / Open, Classified
Recall record
- Recall number
- Z-0175-2027
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medline Industries, LP
- Product
- Medline surgical convenience kits labeled as: 1) CV BASIN PACK-LF, Medline Product number/SKU DYNJ0538615K; 2) CV BASIN PACK-LF, Medline product number/SKU DYNJ0538615J; 3) CV BASIN PACK-LF, Medline product number/SKU DYNJ0538615L.
- Product code
- KDD Kit, Surgical Instrument, Disposable
- Reason for recall
- Medical convenience kits were distributed containing cannula components which do not conform to ISO 80369-7 dimensional requirements and may contribute to leaking or breaking during use.
- FDA root cause
- Under Investigation by firm
- Action
- Medline Industries issued a MEDICAL DEVICE RECALL notice to its consignees on 9/2/2026 via USPS first class mail and email. The notice explained the issue, potential risk, and requested the following: "REQUIRED ACTION: 1.Immediately check your stock for the affected item number(s) and the affected lot number(s) listed within the recall portal. Quarantine all affected product. 2.Please use the link and the information below to complete your response form. Please list the quantity of affected product you have in inventory on the form. Even if you do not have any affected product in inventory, please complete and submit the form. Website link: https://recalls.medline.com, Recall Reference #: R-26-201-FG, Recall Code: 3.Upon receipt of your submitted response form, your account will receive over-labels to place on affected inventory, with instructions for your staff to remove and discard the affected component prior to using the kit. These labels will be provided via FedEx overnight, if applicable." Those consignees who further distributed the product were directed to notify the parties to whom the affected product was distributed. For questions, contact the Recall Department at 866-359-1704 or [email protected].
- Quantity in commerce
- 235 kits
- Distribution
- US Nationwide distribution.
- Product codes and lots
- Medline Product Number/SKU DYNJ0538615L; UDI/DI 10198459414152 (each) 40198459414153 (case), Lot number 26BBA206; Medline Product Number/SKU DYNJ0538615L; UDI/DI 10198459414152 (each) 40198459414153 (case), Lot number 25LBQ023; Medline Product Number/SKU DYNJ0538615L; UDI/DI 10198459414152 (each) 40198459414153 (case), Lot number 25KBR149; Medline Product Number/SKU DYNJ0538615K; UDI/DI 10198459136429 (each) 40198459136420 (case), Lot number 25EBF588; Medline Product Number/SKU DYNJ0538615K; UDI/DI 10198459136429 (each) 40198459136420 (case), Lot number 25ABI296; Medline Product Number/SKU DYNJ0538615K; UDI/DI 10198459136429 (each) 40198459136420 (case), Lot number 24IBP190; Medline Product Number/SKU DYNJ0538615J; UDI/DI 10195327673376 (each) 40195327673377 (case), Lot number 24DBC864.