Z-0182-2027Class II
Convenience Kits containing CONMED AirSeal Access Ports GENERAL DAVINCI DYNJG901004C LAP CHOLE DYNJ88071...
Medline Industries, LP / initiated 2026-08-31 / Open, Classified
Recall record
- Recall number
- Z-0182-2027
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medline Industries, LP
- Product
- Convenience Kits containing CONMED AirSeal Access Ports GENERAL DAVINCI DYNJG901004C LAP CHOLE DYNJ88071 ROBOTIC PACK DYNJ50474F
- Product code
- LRO General Surgery Tray
- Reason for recall
- Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
- FDA root cause
- Process change control
- Action
- The firm began notifying customers on August 31, 2026. Customers were provided with CONMED's original recall notification, which instructs users to have a replacement AirSeal disposable be available prior to the procedure in order to minimize potential delays. Users should ensure tubing is properly connected between the tube set and access ports prior to initiating insufflation. If the connection cannot be securely tightened and the "overpressure" alert cannot be resolved, users should replace the affected device. Users were also provided with instructions on how to connect the AirSeal tubeset and access ports, which include new figures as an update to the prior Instructions for Use. The corrected Instructions for Use are available electronically at eifu.conmed.com/eifu.
- Quantity in commerce
- 3,551 kits
- Distribution
- US Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI.
- Product codes and lots
- DYNJG901004C UDI-DI 10198459535772 Lots 25LLA307 26BLA900 DYNJ88071 UDI-DI 10198459098512 Lots 24KMD416 24LMG268 25AMH987 25BMA076 25GMK411 DYNJ50474F UDI-DI 10195327559007 Lots 24DDA107 24DDA469 24EDA512 24GDB884 24IDA882
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0180-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0181-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0182-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0183-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0184-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0185-2027 | Class II | Ongoing | Voluntary: Firm initiated |