Z-0185-2027Class II
Convenience Kits containing CONMED AirSeal Access Ports KIT ROBOTICS UROLOGY PROSTATE DYKMBNDL116G
Medline Industries, LP / initiated 2026-08-31 / Open, Classified
Recall record
- Recall number
- Z-0185-2027
- Classification
- Class II
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Medline Industries, LP
- Product
- Convenience Kits containing CONMED AirSeal Access Ports KIT ROBOTICS UROLOGY PROSTATE DYKMBNDL116G
- Product code
- OKV Obstetrical Kit
- Reason for recall
- Kits contain CONMED AirSeal Access Ports. The ports were recalled by manufacturer due to potential for erroneous overpressure alerts and system venting during use, which may result in delays or ending procedures.
- FDA root cause
- Process change control
- Action
- The firm began notifying customers on August 31, 2026. Customers were provided with CONMED's original recall notification, which instructs users to have a replacement AirSeal disposable be available prior to the procedure in order to minimize potential delays. Users should ensure tubing is properly connected between the tube set and access ports prior to initiating insufflation. If the connection cannot be securely tightened and the "overpressure" alert cannot be resolved, users should replace the affected device. Users were also provided with instructions on how to connect the AirSeal tubeset and access ports, which include new figures as an update to the prior Instructions for Use. The corrected Instructions for Use are available electronically at eifu.conmed.com/eifu.
- Quantity in commerce
- 3,551 kits
- Distribution
- US Nationwide distribution in the states of AZ, CA, IN, MD, MO, NJ, OH, and WI.
- Product codes and lots
- DYKMBNDL116G UDI-DI 10198459156991 Lots 24KBD429 24LBE008 24LBH410 25ABF361 25BBI842 25EBM426 25GBG086 25JBF391 25KBA114 25KBN017 25LBQ601
Enforcement reports
One row per recalled product line| Recall number | Class | Status | Voluntary or mandated | Report date | Classified |
|---|---|---|---|---|---|
| Z-0180-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0181-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0182-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0183-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0184-2027 | Class II | Ongoing | Voluntary: Firm initiated | ||
| Z-0185-2027 | Class II | Ongoing | Voluntary: Firm initiated |