Z-0289-2027
Alphenix Interventional Angiography System Ref: Infx-8000C
Canon Medical System, USA, INC. / initiated 2026-10-12 / Open, Classified
Recall record
- Recall number
- Z-0289-2027
- Status
- Open, Classified
- Initiated
- Posted
- Recalling firm
- Canon Medical System, USA, INC.
- Product
- Alphenix Interventional Angiography System Ref: Infx-8000C
- Product code
- OWB Interventional Fluoroscopic X-Ray System
- Reason for recall
- Due to software issue, X-ray interventional system may experience internal error if F-Store or imagine acquisition is performed while clinical analysis function is use.
- FDA root cause
- Software design
- Action
- On October 12, 2026, Canon Medical Systems USA, Inc. issued a Urgent: Medical Device Correction to affected consignees. Canon provided the following to consignees: 1. A representative of Canon Medical Systems USA will contact you to schedule a time to conduct installation of modified software to prevent internal errors from occuring if F-STORE or image acquisition is performed while CAAS clinical analysis function is used. 2. If you have any questions regarding this matter, please contact your service representative or contact our InTouch team at (800) 521-1968 or by email [email protected]. 3. In order to comply with the FDA record keeping requirement, Canon respectfully requests that the attached form be completed and returned via email to [email protected]
- Quantity in commerce
- 15 SYSTEMS
- Distribution
- US Nationwide distribution in the states of AL, AR, AZ, CA, FL, GA, HI, IL, LA, MA, MI, MN, MS, NC, NV, OH, OR, TX.
- Product codes and lots
- DI-04987670100192