Z-0301-2027Class II

EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation...

BioMerieux SA / initiated 2026-09-11 / Open, Classified

Recall record

Recall number
Z-0301-2027
Classification
Class II
Status
Open, Classified
Initiated
Posted
Recalling firm
BioMerieux SA
Product
EMAG system Model/Catalog Number: 418591/418591U Product Description: Nucleic acid sample preparation instrument IVD, automated
Product code
JJH Clinical Sample Concentrator
Reason for recall
Delayed or no result when analyzing clinical samples with EMAG system (ref: 418591/418591U) due to a randomly occurring software error, Error 10003 Heater timeout after updating to recent firmware SP4 Patch 2
FDA root cause
Software change control
Action
The firm began notifying consignees on September 11, 2026. Customers were informed they may continue use of EMAG systems and there is no risk of false results. In case of 10003 error heater time-out during an extraction protocol, customers are instructed to reset the system to run the protocol again to prevent delay/loss of sample. The company will contact customers to schedule a bioMerieux Field Service Engineer (FSE) visit to update affected EMAG systems to the previous firmware patch SP4 Patch 1 to fix the issue.
Quantity in commerce
33
Distribution
Worldwide - US Nationwide distribution in the states of CA, IN, KS, LA, MN, MO, NY, OH, and TX. The countries of Canada, United Kingdom, Sweden, France, Norway, Netherlands, Belgium, Denmark, Germany, Austria.
Product codes and lots
UDI-DI 357302BUDI000544TP Serial Numbers IM03343 IM03283 IM03104 IM03329 IM03376 IM03474 IM03321 IM03071 IM03175 IM03194 IM03515 IM03516 IM03240 IM03478 IM03479 IM03482 IM03034 IM03092 IM03378 IM03414 IM03177 IM03193 IM03506 IM3054 IM03439 IM03350 IM03380 IM03131 IM03148 IM03153 IM03275 IM03314 IM03351

Enforcement reports

One row per recalled product line
FDA enforcement reports joined to recall event Z-0301-2027
Recall numberClassStatusVoluntary or mandatedReport dateClassified
Z-0301-2027Class IIOngoingVoluntary: Firm initiated